Using real-time nuclear activation detectors for measuring neutron yields from D(D, T)n reactions on the national ignition facility (NIF)
The National Ignition Facility (NIF) has 48 Real-Time Nuclear Activation Detectors distributed around the target chamber capable of measuring deuterium-triton reaction neutron yields with high precision. In this work, we extend this functionality to deuterium-deuterium (DD) reaction neutrons using a nuclear reaction that occurs in the detector's scintillator material. The corresponding decay of the activated material has a very short half-life of 5 s, which necessitates rapid data collection immediately following an experiment. In this regime, dead time can be very high (>50%) adding significant uncertainty to the measurement. To combat this, we have developed a dead time model that can self-consistently describe the measured data. Initial results show reasonable agreement (within 20%) with DD neutron yields from neutron time-of-flight spectrometers.
Authors
- D. J. Schlossberg (ORCID: https://orcid.org/0000-0002-8713-9448)
- Kelly Hahn
- F. M. Munteanu
- E. A. Henry
- B. Lahmann (ORCID: https://orcid.org/0000-0002-4599-8463)
- R. M. Bionta
Institutions
- Lawrence Livermore National Laboratory (US)
Publication Details
- Journal
- Review of Scientific Instruments
- Published
- 2025-03-01
- DOI
- https://doi.org/10.1063/5.0213464
- Primary Topic
- Nuclear Physics and Applications
- Type
- article
- Field-Weighted Citation Impact
- 0.00
Funders
- U.S. Department of Energy