Importance of Raw‐Data Analysis When Determining the Cause of Pause Alerts Transmitted by an Implantable Loop Recorder Remote Monitoring Devices
ILR RM pause alerts should always be reviewed via the raw-data RM web page or alternatively via ILR device interrogations, to achieve a sufficiently high-resolution image to assure these pauses are not due to AVB.
Authors
- Michael Glikson (ORCID: https://orcid.org/0000-0001-5706-0395)
- Moshe Rav (ORCID: https://orcid.org/0000-0002-5365-9947)
- Michael Ilan
- Hana Raiz
Institutions
- Hebrew University of Jerusalem (IL)
- Shaare Zedek Medical Center (IL)
Publication Details
- Journal
- Journal of Cardiovascular Electrophysiology
- Published
- 2025-03-06
- DOI
- https://doi.org/10.1111/jce.16634
- Primary Topic
- Cardiovascular Syncope and Autonomic Disorders
- Type
- article
- Field-Weighted Citation Impact