V‐<scp>PRO</scp> Blood Collection Tubes: Validation for Clinical Chemistry and Immunoassay Tests
Laboratories currently using BD vacutainers should exercise caution if they intend to perform chemistry and immunoassay tests with V-PRO tubes. The technical validation outcomes for V-PRO were not acceptable due to significant faults identified in comparison to BD Vacutainer.