Accelerometer-based estimation of respiratory rate using principal component analysis and autocorrelation

<i>Objective.</i>Respiratory rate (RR) is an important vital sign but is often neglected. Multiple technologies exist for RR monitoring but are either expensive or impractical. Tri-axial accelerometry represents a minimally intrusive solution for continuous RR monitoring, however, the method has not been validated in a wide RR range. Therefore, the aim of this study was to investigate the agreement between RR estimation from a tri-axial accelerometer and a reference method in a wide RR range.<i>Approach.</i>Twenty-five healthy participants were recruited. For accelerometer RR estimation, the accelerometer was placed on the abdomen for optimal breathing movement detection. The acquired accelerometry data were processed using a lowpass filter, principal component analysis (PCA), and autocorrelation. The subjects were instructed to breathe at slow, normal, and fast paces in segments of 60 s. A flow meter was used as reference. Furthermore, the PCA-autocorrelation method was compared with a similar single axis method.<i>Main results.</i>The PCA-autocorrelation method resulted in a bias of 0.0 breaths per minute (bpm) and limits of agreement (LOA) = [-1.9; 1.9 bpm] compared to the reference. Overall, 99% of the RRs estimated by the PCA-autocorrelation method were within ±2 bpm of the reference. A Pearson correlation indicated a very strong correlation with<i>r</i> = 0.99 (p<0.001). The single axis method resulted in a bias of 3.7 bpm, LOA = [-14.9; 22.3 bpm], and<i>r</i> = 0.44 (p<0.001).<i>Significance.</i>The results indicate a strong agreement between the PCA-autocorrelation method and the reference. Furthermore, the PCA-autocorrelation method outperformed the single axis method.

Authors

Institutions

Publication Details

Journal
Physiological Measurement
Published
2025-03-07
DOI
https://doi.org/10.1088/1361-6579/adbe23
Primary Topic
Non-Invasive Vital Sign Monitoring
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Accelerometer-based estimation of respiratory rate using principal component analysis and autocorrelation

Samuel Emil Schmidt, Dan Stieper Karbing, Anne Sofie Nielsen, Emil Korsgaard et al.
Physiological Measurement
Non-Invasive Vital Sign Monitoring
article

Accelerometer-based estimation of respiratory rate using principal component analysis and autocorrelation

Samuel Emil Schmidt, Dan Stieper Karbing, Anne Sofie Nielsen, Emil Korsgaard, Mads C F Hostrup, Freja E Sørensen, Jesper O Kragballe, Morten U Østergaard
article en

Abstract

<i>Objective.</i>Respiratory rate (RR) is an important vital sign but is often neglected. Multiple technologies exist for RR monitoring but are either expensive or impractical. Tri-axial accelerometry represents a minimally intrusive solution for continuous RR monitoring, however, the method has not been validated in a wide RR range. Therefore, the aim of this study was to investigate the agreement between RR estimation from a tri-axial accelerometer and a reference method in a wide RR range.<i>Approach.</i>Twenty-five healthy participants were recruited. For accelerometer RR estimation, the accelerometer was placed on the abdomen for optimal breathing movement detection. The acquired accelerometry data were processed using a lowpass filter, principal component analysis (PCA), and autocorrelation. The subjects were instructed to breathe at slow, normal, and fast paces in segments of 60 s. A flow meter was used as reference. Furthermore, the PCA-autocorrelation method was compared with a similar single axis method.<i>Main results.</i>The PCA-autocorrelation method resulted in a bias of 0.0 breaths per minute (bpm) and limits of agreement (LOA) = [-1.9; 1.9 bpm] compared to the reference. Overall, 99% of the RRs estimated by the PCA-autocorrelation method were within ±2 bpm of the reference. A Pearson correlation indicated a very strong correlation with<i>r</i> = 0.99 (p<0.001). The single axis method resulted in a bias of 3.7 bpm, LOA = [-14.9; 22.3 bpm], and<i>r</i> = 0.44 (p<0.001).<i>Significance.</i>The results indicate a strong agreement between the PCA-autocorrelation method and the reference. Furthermore, the PCA-autocorrelation method outperformed the single axis method.

Physiological MeasurementVol. 46(3)
Aalborg University (DK)
Openalex Percentile: Top 96%
Non-Invasive Vital Sign Monitoring
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Accelerometer-based estimation of respiratory rate using principal component analysis and autocorrelation — Samuel Emil Schmidt, Dan Stieper Karbing, et al. · Physiological Measurement (2025) | TGRS Research Map | TGRS