Influence of Measurement Geometry and Blank on Absolute Measurements of Photoluminescence Quantum Yields of Scattering Luminescent Films

For a series of 500 μm-thick polyurethane films containing different concentrations of luminescent and scattering YAG:Ce microparticles, we systematically explored and quantified pitfalls of absolute measurements of photoluminescence quantum yields (Φ<sub>f</sub>) for often employed integrating sphere (IS) geometries, where the sample is placed either on a sample holder at the bottom of the IS surface or mounted in the IS center. Thereby, the influence of detection and illumination geometry and sample position was examined using blanks with various scattering properties for measuring the number of photons absorbed by the sample. Our results reveal that (i) setup configurations where the scattering sample is mounted in the IS center and (ii) transparent blanks can introduce systematic errors in absolute Φ<sub>f</sub> measurements. For strongly scattering, luminescent samples, this can result in either an under- or overestimation of the absorbed photon flux and hence an under- or overestimation of Φ<sub>f</sub>. The size of these uncertainties depends on the scattering properties of the sample and instrument parameters, such as sample position, IS size, wavelength-dependent reflectivity of the IS surface coating, and port configuration. For accurate and reliable absolute Φ<sub>f</sub> measurements, we recommend (i) a blank with scattering properties closely matching those of the sample to realize similar distributions of the diffusely scattered excitation photons within the IS, and (ii) a sufficiently high sample absorption at the excitation wavelength. For IS setups with center-mounted samples, measurement geometries should be utilized that prevent the loss of excitation photons by reflections from the sample out of the IS.

Authors

Institutions

Publication Details

Journal
Analytical Chemistry
Published
2025-03-07
DOI
https://doi.org/10.1021/acs.analchem.4c06726
Primary Topic
Luminescence Properties of Advanced Materials
Type
article
Field-Weighted Citation Impact
0.00

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Influence of Measurement Geometry and Blank on Absolute Measurements of Photoluminescence Quantum Yields of Scattering Luminescent Films

Florian Frenzel, Christian Würth, Ute Resch‐Genger, Arne Güttler et al.
Analytical Chemistry
Luminescence Properties of Advanced Materials
article

Influence of Measurement Geometry and Blank on Absolute Measurements of Photoluminescence Quantum Yields of Scattering Luminescent Films

Florian Frenzel, Christian Würth, Ute Resch‐Genger, Arne Güttler, Saskia Fiedler, Ahmad Bardan
article en

Abstract

For a series of 500 μm-thick polyurethane films containing different concentrations of luminescent and scattering YAG:Ce microparticles, we systematically explored and quantified pitfalls of absolute measurements of photoluminescence quantum yields (Φ<sub>f</sub>) for often employed integrating sphere (IS) geometries, where the sample is placed either on a sample holder at the bottom of the IS surface or mounted in the IS center. Thereby, the influence of detection and illumination geometry and sample position was examined using blanks with various scattering properties for measuring the number of photons absorbed by the sample. Our results reveal that (i) setup configurations where the scattering sample is mounted in the IS center and (ii) transparent blanks can introduce systematic errors in absolute Φ<sub>f</sub> measurements. For strongly scattering, luminescent samples, this can result in either an under- or overestimation of the absorbed photon flux and hence an under- or overestimation of Φ<sub>f</sub>. The size of these uncertainties depends on the scattering properties of the sample and instrument parameters, such as sample position, IS size, wavelength-dependent reflectivity of the IS surface coating, and port configuration. For accurate and reliable absolute Φ<sub>f</sub> measurements, we recommend (i) a blank with scattering properties closely matching those of the sample to realize similar distributions of the diffusely scattered excitation photons within the IS, and (ii) a sufficiently high sample absorption at the excitation wavelength. For IS setups with center-mounted samples, measurement geometries should be utilized that prevent the loss of excitation photons by reflections from the sample out of the IS.

Analytical ChemistryVol. 97(10)
Federal Institute For Materials Research and Testing (DE)
Deutsche Forschungsgemeinschaft
Openalex Percentile: Top 97%
Luminescence Properties of Advanced Materials
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.