Calibrating p-i-n Diodes for Displacement Damage Monitoring in a Space Radiation Environment

This work investigates the response of specially developed silicon p-i-n diodes with a long base as sensors to measure displacement damage dose (DDD) in silicon. Measurements of DDD are based on the forward voltage shift (<inline-formula xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink"> <tex-math notation="LaTeX">$\Delta V_{F}$ </tex-math></inline-formula>) of irradiated p-i-n diodes. In this work, the p-i-n diodes were irradiated in the following radiation fields: 1) 20-MeV electrons from a medical linear accelerator (LINAC); 2) 200-, 150-, and 100-MeV protons at the Groningen proton therapy facility; and 3) 4.63- and 15.3-MeV quasi-monoenergetic neutrons at the Czech Technical University, Prague. It was demonstrated that the calibration factor (<inline-formula xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink"> <tex-math notation="LaTeX">$\alpha $ </tex-math></inline-formula>), which is the response of the p-i-n diode in terms of DDD obtained using 20-MeV electrons, can be used to predict the p-i-n diode response for protons and neutrons as mentioned above (within experimental error). However, the p-i-n diode response related to 100-MeV protons has a 70% higher response than predicted, which requires further investigation. Overall, the 20-MeV electrons from a medical LINAC were found to be suitable for p-i-n diodes calibration in terms of the non-ionizing energy losses (NIELs) and are widely available in contrast to 1-MeV electrons, fast neutrons, and protons.

Authors

Institutions

Publication Details

Journal
IEEE Transactions on Nuclear Science
Published
2025-03-07
DOI
https://doi.org/10.1109/tns.2025.3549051
Citations
1
Primary Topic
Sensor Technology and Measurement Systems
Type
article
Field-Weighted Citation Impact
1.49

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Calibrating p-i-n Diodes for Displacement Damage Monitoring in a Space Radiation Environment

James Vohradsky, I.E. Anokhin, Joel Poder, Dawn J. Bennett et al.
1 citations
IEEE Transactions on Nuclear Science
Sensor Technology and Measurement Systems
1.49
article

Calibrating p-i-n Diodes for Displacement Damage Monitoring in a Space Radiation Environment

James Vohradsky, I.E. Anokhin, Joel Poder, Dawn J. Bennett, R.E. Sykora, Vladimir Pan, Anatoly Rosenfeld, A. Perevertaylo, Z. Kohout, Hugo Cintas, Linh T. Tran, B. Bergmann, Dirk Wagenaar, T. Slavíček, P. Smolyankiy
article en
1 citations

Abstract

This work investigates the response of specially developed silicon p-i-n diodes with a long base as sensors to measure displacement damage dose (DDD) in silicon. Measurements of DDD are based on the forward voltage shift (<inline-formula xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink"> <tex-math notation="LaTeX">$\Delta V_{F}$ </tex-math></inline-formula>) of irradiated p-i-n diodes. In this work, the p-i-n diodes were irradiated in the following radiation fields: 1) 20-MeV electrons from a medical linear accelerator (LINAC); 2) 200-, 150-, and 100-MeV protons at the Groningen proton therapy facility; and 3) 4.63- and 15.3-MeV quasi-monoenergetic neutrons at the Czech Technical University, Prague. It was demonstrated that the calibration factor (<inline-formula xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink"> <tex-math notation="LaTeX">$\alpha $ </tex-math></inline-formula>), which is the response of the p-i-n diode in terms of DDD obtained using 20-MeV electrons, can be used to predict the p-i-n diode response for protons and neutrons as mentioned above (within experimental error). However, the p-i-n diode response related to 100-MeV protons has a 70% higher response than predicted, which requires further investigation. Overall, the 20-MeV electrons from a medical LINAC were found to be suitable for p-i-n diodes calibration in terms of the non-ionizing energy losses (NIELs) and are widely available in contrast to 1-MeV electrons, fast neutrons, and protons.

IEEE Transactions on Nuclear ScienceVol. 72(4)
University Medical Center Groningen (NL), Kyiv National University of Construction and Architecture (UA), University of Groningen (NL), University of Wollongong (AU), Dialyse Centrum Groningen (NL), Institute for Nuclear Research (UA), Czech Technical University in Prague (CZ)
Australian Research Council
Climate action
Openalex Percentile: Top 19%
Sensor Technology and Measurement Systems
1.49
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.