Physiological Scaffold Remodeling in the Coronary Artery After 30 Months of Bioadaptor Implantation
The DynamX Bioadaptor (Elixir Medical Corporation) is a novel coronary artery scaffold consisting of 3 helical strands connected by a bioabsorbable polymer. After 6 months, the polymer dissolves, restoring the physiological motion of the coronary artery as the interconnected (caged) structure of the scaffold disengages. Here we report the cases of 2 patients treated with the Bioadaptor who underwent intravascular imaging at the time of the index procedure and during follow-up periods of 30 and 36 months, respectively. The cross-sectional images of the scaffold were analyzed and compared with those obtained during the index procedure. At follow-up, the mean cross-sectional area of the scaffold was increased significantly in both cases, regardless of the degree of intimal hyperplasia. Such unique properties of the Bioadaptor make it a promising alternative to contemporary drug-eluting stents.
Authors
- Shiori Kawakami
- Takeshi Yamada (ORCID: https://orcid.org/0000-0002-6490-758X)
- Akihiko Takahashi (ORCID: https://orcid.org/0000-0003-4906-2328)
- Shunsuke Nakajima
- Norimasa Taniguchi
- Shigeru Saito (ORCID: https://orcid.org/0000-0002-5985-4815)
- Tetsuya Hata
Institutions
- Kobe Asahi Hospital (JP)
- Shonan Kamakura General Hospital (JP)
Publication Details
- Journal
- JACC Case Reports
- Published
- 2025-03-01
- DOI
- https://doi.org/10.1016/j.jaccas.2024.103089
- Citations
- 3
- Primary Topic
- Coronary Interventions and Diagnostics
- Type
- article
- Field-Weighted Citation Impact
- 6.02