Investigation of optical wireless power transmission across biological tissue using NIR light considering safe irradiance level

Presently, implantable electronic devices (IEDs) are becoming integral choices in medical applications, still their reliance on nonrechargeable batteries limits their longevity. This challenge has driven worldwide interest in developing rechargeable IEDs. While energy transfer based on radio-frequency (RF) has been clinically employed to address the challenge of IEDs running out of battery power, research into alternative methods, such as near-infrared (NIR) light-based energy transfer schemes, is advancing rapidly. The key advantages of optical systems are security, privacy, safety, interference-free operation, etc. The properties of NIR light are attractive for optical wireless power transmission (OWPT) across biological tissue due to minimal scattering and absorption effect. NIR light has been widely employed for various therapeutic applications, such as photobiomodulation (PBM) therapy. When using power transfer, ensuring that do not exceed safety regulations to avoid damage to the tissue is important. In this study, we investigate OWPT under safe level irradiance referring to the typical PBM applications, i.e., 20 mW/cm 2 , 25 mW/cm 2 , and 200 mW/cm 2 . The OWPT system is tested on an experimental testbed, consisting of an 810 nm NIR LED and a photovoltaic (PV) cell as transmitter and receiver, respectively. A tissue-mimicking optical phantom was employed as a propagation media that has long-lasting stability at a room temperature around the testbed, thus suitable for prolonged experiments. We measure the time it takes to charge an energy-storing supercapacitor under three different safe irradiance levels to assess the feasibility of safely charging an IED’s storage. In addition, we compare the results with the maximum irradiance of NIR LED, i.e., 527 mW/cm 2 . Comparing 527 mW/cm 2 to PBM irradiance revealed a trade-off between supercapacitor charging time and tissue safety. Charging limitations and other related issues are also elaborated in this paper

Authors

Institutions

Publication Details

Journal
Results in Optics
Published
2025-03-05
DOI
https://doi.org/10.1016/j.rio.2025.100801
Citations
2
Primary Topic
Molecular Communication and Nanonetworks
Type
article
Field-Weighted Citation Impact
1.35
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Investigation of optical wireless power transmission across biological tissue using NIR light considering safe irradiance level

Mariella Särestöniemi, Syifaul Fuada, Marcos Katz
2 citations
Results in Optics
Molecular Communication and Nanonetworks
1.35
article

Investigation of optical wireless power transmission across biological tissue using NIR light considering safe irradiance level

Mariella Särestöniemi, Syifaul Fuada, Marcos Katz
article en
2 citations

Abstract

Presently, implantable electronic devices (IEDs) are becoming integral choices in medical applications, still their reliance on nonrechargeable batteries limits their longevity. This challenge has driven worldwide interest in developing rechargeable IEDs. While energy transfer based on radio-frequency (RF) has been clinically employed to address the challenge of IEDs running out of battery power, research into alternative methods, such as near-infrared (NIR) light-based energy transfer schemes, is advancing rapidly. The key advantages of optical systems are security, privacy, safety, interference-free operation, etc. The properties of NIR light are attractive for optical wireless power transmission (OWPT) across biological tissue due to minimal scattering and absorption effect. NIR light has been widely employed for various therapeutic applications, such as photobiomodulation (PBM) therapy. When using power transfer, ensuring that do not exceed safety regulations to avoid damage to the tissue is important. In this study, we investigate OWPT under safe level irradiance referring to the typical PBM applications, i.e., 20 mW/cm 2 , 25 mW/cm 2 , and 200 mW/cm 2 . The OWPT system is tested on an experimental testbed, consisting of an 810 nm NIR LED and a photovoltaic (PV) cell as transmitter and receiver, respectively. A tissue-mimicking optical phantom was employed as a propagation media that has long-lasting stability at a room temperature around the testbed, thus suitable for prolonged experiments. We measure the time it takes to charge an energy-storing supercapacitor under three different safe irradiance levels to assess the feasibility of safely charging an IED’s storage. In addition, we compare the results with the maximum irradiance of NIR LED, i.e., 527 mW/cm 2 . Comparing 527 mW/cm 2 to PBM irradiance revealed a trade-off between supercapacitor charging time and tissue safety. Charging limitations and other related issues are also elaborated in this paper

Results in OpticsVol. 19
University of Oulu (FI)
Affordable and clean energy
Openalex Percentile: Top 22%
Molecular Communication and Nanonetworks
1.35
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.