Effect of Variation Number of Excitation (NEX) Parameter on the Quality Image of Magnetic Resonance Imaging Genu on T2 Weighted Image Sequence at Bali Mandara Regional General Hospital, Indonesia

Aim: Study on the effect of NEX parameter variations on image quality in T2WI sequences on genu MRI images aims to determine the effect of Number of Excitation (NEX) variations on image quality which is influenced by scanning time and Signal to Noise Ratio (SNR) and determine the best NEX variation to produce scanning time and SNR T2WI MRI genu. Study Design: This study used a quantitative analytic design in the form of experiments and direct field observations. This study was conducted in the MRI unit of Bali Mandara Hospital from January to December 2024. Methodology: This study involved 30 patients who underwent genu MRI examination with 3 variations of NEX values ranging from (1, 2 and 3). Evaluating the scanning time and SNR value of the tissue by measuring the ROI directly on the MRI device. Data were analyzed using IBM SPSS statistical software and Excel software to evaluate the effect of NEX variation on image quality. The MRI machine used was Philips Ingenia CX 1.5 Tesla with standard protocol parameters for imaging genu Results: Scanning time and SNR values were analyzed on ligament, bone and fat tissues. The time required for scanning time was 115 seconds, 230 seconds and 345 seconds, respectively. Sequentially, the average SNR values obtained based on the calculations performed were 6.906, 10.203 and 13.307 in ligament tissue, 81.209, 115.289 and 136.14 in bone tissue, and 106.662, 151.3 and 178.212 in fat tissue. Increasing the NEX value significantly increased the scanning time and SNR value. NEX 3 provides the highest scanning time and SNR value compared to other NEX variations. This indicates that multiple NEXs produce better image quality. Conclusion: Good MRI image quality is determined by high scanning time and SNR values. This study recommends the use of NEX 3 in genu MRI imaging protocols to improve the accuracy of clinical examinations.

Authors

Institutions

Publication Details

Journal
Asian Journal of Medicine and Health
Published
2025-03-06
DOI
https://doi.org/10.9734/ajmah/2025/v23i41199
Primary Topic
Radiomics and Machine Learning in Medical Imaging
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Effect of Variation Number of Excitation (NEX) Parameter on the Quality Image of Magnetic Resonance Imaging Genu on T2 Weighted Image Sequence at Bali Mandara Regional General Hospital, Indonesia

Ni Luh Putu Trisnawati, Anak Agung Ngurah Gunawan, Ni Kadek Nova Anggarani, I Gusti Agung Widagda et al.
Asian Journal of Medicine and Health
Radiomics and Machine Learning in Medical Imaging
article

Effect of Variation Number of Excitation (NEX) Parameter on the Quality Image of Magnetic Resonance Imaging Genu on T2 Weighted Image Sequence at Bali Mandara Regional General Hospital, Indonesia

Ni Luh Putu Trisnawati, Anak Agung Ngurah Gunawan, Ni Kadek Nova Anggarani, I Gusti Agung Widagda, Komang Ngurah Suarbawa, Ni Nyoman Ratini, Ni Putu Sri Wulan Ningsih
article en

Abstract

Aim: Study on the effect of NEX parameter variations on image quality in T2WI sequences on genu MRI images aims to determine the effect of Number of Excitation (NEX) variations on image quality which is influenced by scanning time and Signal to Noise Ratio (SNR) and determine the best NEX variation to produce scanning time and SNR T2WI MRI genu. Study Design: This study used a quantitative analytic design in the form of experiments and direct field observations. This study was conducted in the MRI unit of Bali Mandara Hospital from January to December 2024. Methodology: This study involved 30 patients who underwent genu MRI examination with 3 variations of NEX values ranging from (1, 2 and 3). Evaluating the scanning time and SNR value of the tissue by measuring the ROI directly on the MRI device. Data were analyzed using IBM SPSS statistical software and Excel software to evaluate the effect of NEX variation on image quality. The MRI machine used was Philips Ingenia CX 1.5 Tesla with standard protocol parameters for imaging genu Results: Scanning time and SNR values were analyzed on ligament, bone and fat tissues. The time required for scanning time was 115 seconds, 230 seconds and 345 seconds, respectively. Sequentially, the average SNR values obtained based on the calculations performed were 6.906, 10.203 and 13.307 in ligament tissue, 81.209, 115.289 and 136.14 in bone tissue, and 106.662, 151.3 and 178.212 in fat tissue. Increasing the NEX value significantly increased the scanning time and SNR value. NEX 3 provides the highest scanning time and SNR value compared to other NEX variations. This indicates that multiple NEXs produce better image quality. Conclusion: Good MRI image quality is determined by high scanning time and SNR values. This study recommends the use of NEX 3 in genu MRI imaging protocols to improve the accuracy of clinical examinations.

Asian Journal of Medicine and HealthVol. 23(4)
Udayana University (ID)
Openalex Percentile: Top 94%
Radiomics and Machine Learning in Medical Imaging
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.