Augmenting the Sensitivity of an SPR Biosensor for Detecting Urine Glucose by Utilizing BaTiO3 and BlueP/WSe2

Using the transfer matrix method (TMM), a hybrid SPR-based biosensor was examined to detect glucose concentration in urine. The Kretschmann configuration with a BK7 glass prism, Ag layer, BaTiO3 dielectric layer, and a 2D BlueP/WSe2 layer was utilized to design the proposed biosensor. The optimized thickness of Ag, BaTiO3, and the monolayer of BlueP/WSe2 were investigated as 52 nm, 11 nm, and 0.34 nm respectively. The layers thickness affects the sensor's performance. In order to achieve the lowest reflectance and maximum sensitivity, the layer's thickness was tuned. The highest sensitivity of 425 deg/RIU, detection accuracy of 0.177305 deg-1, FWHM of 5.64 deg, and quality factor of 75.35 RIU-1 were obtained at a glucose concentration of 10 g/dL.

Authors

Institutions

Publication Details

Published
2024-10-17
DOI
https://doi.org/10.1109/atc63255.2024.10908171
Primary Topic
Electrochemical sensors and biosensors
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Augmenting the Sensitivity of an SPR Biosensor for Detecting Urine Glucose by Utilizing BaTiO3 and BlueP/WSe2

Nguyen Hoang Hai, Md. Anwar Hossain, Abbas F. Jawad, Momen Sahriar Shoshi
Electrochemical sensors and biosensors
article

Augmenting the Sensitivity of an SPR Biosensor for Detecting Urine Glucose by Utilizing BaTiO3 and BlueP/WSe2

Nguyen Hoang Hai, Md. Anwar Hossain, Abbas F. Jawad, Momen Sahriar Shoshi
article en

Abstract

Using the transfer matrix method (TMM), a hybrid SPR-based biosensor was examined to detect glucose concentration in urine. The Kretschmann configuration with a BK7 glass prism, Ag layer, BaTiO3 dielectric layer, and a 2D BlueP/WSe2 layer was utilized to design the proposed biosensor. The optimized thickness of Ag, BaTiO3, and the monolayer of BlueP/WSe2 were investigated as 52 nm, 11 nm, and 0.34 nm respectively. The layers thickness affects the sensor's performance. In order to achieve the lowest reflectance and maximum sensitivity, the layer's thickness was tuned. The highest sensitivity of 425 deg/RIU, detection accuracy of 0.177305 deg-1, FWHM of 5.64 deg, and quality factor of 75.35 RIU-1 were obtained at a glucose concentration of 10 g/dL.

Bangladesh University of Business and Technology (BD), University of Rajshahi (BD), Hanoi University of Science and Technology (VN)
Zero hunger
Openalex Percentile: Top 74%
Electrochemical sensors and biosensors
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.