A Realistic Breast Phantom for Investigating the Features of the Microwave Radiometry Method Using Mathematical and Physical Modelling

This article presents the development of an anatomical breast phantom for investigating the capabilities of microwave radiometry in assessing thermal processes in biological tissues. The phantom accounts for the heterogeneous tissue structure and haemodynamics, enabling realistic heat transfer modelling. Numerical simulation software was developed, accurately reproducing experimental results and allowing the study of thermal anomalies. Experimental validation demonstrated that the temperature in the subcutaneous layer differed on average by 0.3 °C from deeper tissues, confirming the method’s effectiveness. The presence of a tumour in the model resulted in a local temperature increase of up to 0.77 °C, highlighting the sensitivity of microwave radiometry to tumour-induced thermal anomalies. These findings contribute to enhancing non-invasive techniques for early breast disease detection.

Authors

Institutions

Publication Details

Journal
Technologies
Published
2025-03-06
DOI
https://doi.org/10.3390/technologies13030106
Citations
7
Primary Topic
Infrared Thermography in Medicine
Type
article
Field-Weighted Citation Impact
9.52

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

A Realistic Breast Phantom for Investigating the Features of the Microwave Radiometry Method Using Mathematical and Physical Modelling

Maxim Polyakov, Danila S. Sirotin
7 citations
Technologies
Infrared Thermography in Medicine
9.52
article

A Realistic Breast Phantom for Investigating the Features of the Microwave Radiometry Method Using Mathematical and Physical Modelling

Maxim Polyakov, Danila S. Sirotin
article en
7 citations

Abstract

This article presents the development of an anatomical breast phantom for investigating the capabilities of microwave radiometry in assessing thermal processes in biological tissues. The phantom accounts for the heterogeneous tissue structure and haemodynamics, enabling realistic heat transfer modelling. Numerical simulation software was developed, accurately reproducing experimental results and allowing the study of thermal anomalies. Experimental validation demonstrated that the temperature in the subcutaneous layer differed on average by 0.3 °C from deeper tissues, confirming the method’s effectiveness. The presence of a tumour in the model resulted in a local temperature increase of up to 0.77 °C, highlighting the sensitivity of microwave radiometry to tumour-induced thermal anomalies. These findings contribute to enhancing non-invasive techniques for early breast disease detection.

TechnologiesVol. 13(3)
Volgograd State University (RU)
Russian Science Foundation
Openalex Percentile: Top 2%
Infrared Thermography in Medicine
9.52
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.