A multi-point calibration method for electron probe microanalysis (EPMA) of indium in sphalerite (ZnS)

This article presents a multi-point calibration approach for electron probe microanalysis (EPMA) for the trace element analysis of indium in sphalerite (ZnS). To define a multi-point calibration curve, indium and cadmium-doped ZnS crystals in a concentration range from 0 (blank) to ~ 1500 µg / g were used. The samples were measured with two different analytical settings (25 kV acceleration voltage and 100 nA beam current as well as 7 kV and 200 nA). The figures of merit including, beam stability, lower limit of detection and limit of quantification as well as the reproducibility and precision are assessed. Equally, the line overlap of Cd and chemical shift due to non-matrix matched standards is discussed. The multi-point calibration approach results in a 2-3 times improved analytical precision compared to the classical calibration approach using only one calibration sample, and detection limits down to about 20 µg / g were achieved.

Authors

Institutions

Publication Details

Journal
Scientific Reports
Published
2025-03-06
DOI
https://doi.org/10.1038/s41598-025-91085-x
Primary Topic
Electron and X-Ray Spectroscopy Techniques
Type
article
Field-Weighted Citation Impact
0.00

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

A multi-point calibration method for electron probe microanalysis (EPMA) of indium in sphalerite (ZnS)

Thomas Ulrich, Thomas Schirmer
Scientific Reports
Electron and X-Ray Spectroscopy Techniques
article

A multi-point calibration method for electron probe microanalysis (EPMA) of indium in sphalerite (ZnS)

Thomas Ulrich, Thomas Schirmer
article en

Abstract

This article presents a multi-point calibration approach for electron probe microanalysis (EPMA) for the trace element analysis of indium in sphalerite (ZnS). To define a multi-point calibration curve, indium and cadmium-doped ZnS crystals in a concentration range from 0 (blank) to ~ 1500 µg / g were used. The samples were measured with two different analytical settings (25 kV acceleration voltage and 100 nA beam current as well as 7 kV and 200 nA). The figures of merit including, beam stability, lower limit of detection and limit of quantification as well as the reproducibility and precision are assessed. Equally, the line overlap of Cd and chemical shift due to non-matrix matched standards is discussed. The multi-point calibration approach results in a 2-3 times improved analytical precision compared to the classical calibration approach using only one calibration sample, and detection limits down to about 20 µg / g were achieved.

Scientific ReportsVol. 15(1)
Clausthal University of Technology (DE)
Technische Universität Clausthal
Openalex Percentile: Top 96%
Electron and X-Ray Spectroscopy Techniques
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.