Investigating the Sensitivity of NIPAM Gel Dosimetry in CyberKnife Systems Using MRI-based R2 Mapping

Abstract Purpose The CyberKnife delivers precise radiation with sharp dose gradients challenging to measure. N-isopropylacrylamide (NIPAM) gel phantom can capture 3D dose distributions effectively. This study investigates the feasibility of using NIPAM gel dosimetry combined with magnetic resonance imaging (MRI)-based transverse relaxation rate (R2) mapping to verify dose distributions delivered by the CyberKnife system. Methods A total of 24 gel samples were divided into 8 dose groups (0, 3, 6, 9, 12, 15, 18, and 21 Gy), with three samples per dose group for repeat measurements. The NIPAM polymer gels were irradiated by a CyberKnife ® M6 FIM system with a 6 MV x-ray beam. MR images of the NIPAM gel phantom were made 24 h after irradiation using the 3T MRI scanner. The R2 maps corresponding to dose distribution were analyzed using Matlab software. Results Our results demonstrate that R2 values increase with rising irradiation doses, confirming dose-dependent sensitivity. The dose-response curve exhibits a slope of 0.1688 Gy⁻¹ s⁻¹, reflecting the gel dosimeter’s sensitivity, with a strong correlation (R² = 0.85, p = 0.001). Additionally, R2 values vary with depth, peaking at the 30 mm dose center point (D center ) and then gradually decreasing, indicating maximum dose accumulation at D center followed by a decline with increasing depth. Conclusion This study demonstrates that NIPAM gel dosimetry with MRI-based R2 mapping effectively verifies CyberKnife dose distributions. It exhibits strong dose sensitivity and accurately captures sharp gradients, highlighting its potential as a superior alternative to conventional dosimetry methods.

Authors

Institutions

Publication Details

Journal
Journal of Medical and Biological Engineering
Published
2025-03-07
DOI
https://doi.org/10.1007/s40846-025-00935-6
Citations
2
Primary Topic
Advanced Radiotherapy Techniques
Type
article
Field-Weighted Citation Impact
3.77
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Investigating the Sensitivity of NIPAM Gel Dosimetry in CyberKnife Systems Using MRI-based R2 Mapping

Shih‐Wei Huang, Feng‐Yi Su, Shin‐Lei Peng, Hao-Ting Lan et al.
2 citations
Journal of Medical and Biological Engineering
Advanced Radiotherapy Techniques
3.77
article

Investigating the Sensitivity of NIPAM Gel Dosimetry in CyberKnife Systems Using MRI-based R2 Mapping

Shih‐Wei Huang, Feng‐Yi Su, Shin‐Lei Peng, Hao-Ting Lan, Chun‐Hsu Yao, Hung-Yi Chen, Yu-Hsuan Lin
article en
2 citations

Abstract

Abstract Purpose The CyberKnife delivers precise radiation with sharp dose gradients challenging to measure. N-isopropylacrylamide (NIPAM) gel phantom can capture 3D dose distributions effectively. This study investigates the feasibility of using NIPAM gel dosimetry combined with magnetic resonance imaging (MRI)-based transverse relaxation rate (R2) mapping to verify dose distributions delivered by the CyberKnife system. Methods A total of 24 gel samples were divided into 8 dose groups (0, 3, 6, 9, 12, 15, 18, and 21 Gy), with three samples per dose group for repeat measurements. The NIPAM polymer gels were irradiated by a CyberKnife ® M6 FIM system with a 6 MV x-ray beam. MR images of the NIPAM gel phantom were made 24 h after irradiation using the 3T MRI scanner. The R2 maps corresponding to dose distribution were analyzed using Matlab software. Results Our results demonstrate that R2 values increase with rising irradiation doses, confirming dose-dependent sensitivity. The dose-response curve exhibits a slope of 0.1688 Gy⁻¹ s⁻¹, reflecting the gel dosimeter’s sensitivity, with a strong correlation (R² = 0.85, p = 0.001). Additionally, R2 values vary with depth, peaking at the 30 mm dose center point (D center ) and then gradually decreasing, indicating maximum dose accumulation at D center followed by a decline with increasing depth. Conclusion This study demonstrates that NIPAM gel dosimetry with MRI-based R2 mapping effectively verifies CyberKnife dose distributions. It exhibits strong dose sensitivity and accurately captures sharp gradients, highlighting its potential as a superior alternative to conventional dosimetry methods.

Journal of Medical and Biological EngineeringVol. 45(2)
Asia University (TW), China Medical University (TW), China Medical University Hospital (TW), National Taiwan University Hospital (TW)
Openalex Percentile: Top 9%
Advanced Radiotherapy Techniques
3.77
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.