Coronary artery blockage, the most common cardiovascular problem, is often treated with drug-eluting stents (DES). This study aims to address the main limitation of traditional angioplasty therapy. Thus, designing, fabricating, and analyzing a novel drug-eluting polymeric stent using liquid crystal display (LCD) technology may potentially represent an innovative approach to DES in the near future. Therefore, a poly (lactic acid) (PLA) based 3D-printed stent was designed using SolidWorks software and fabricated using the liquid crystal display (LCD) method. The cyclosporine-loaded stent was prepared and analyzed using optical microscopy, differential scanning calorimetry (DSC), and Fourier transform infrared spectroscopy (FTIR). Loading efficiency percentage and release characteristics were estimated. The polymeric stent platform was successfully designed and manufactured using PLA resin. Based on the characterization of cyclosporine eluting stent, a loading efficiency of 48.66 ± 5.92% was estimated through the immersion method. The FTIR and DSC results confirmed molecular interactions between cyclosporine and the PLA-based 3D-printed stent compared with physical mixture formulations. A sustained release profile of cyclosporine was also observed where approximately 50% of the drug was released within the first three hours. The sustained-release profile, characterized by the absence of a burst release, holds significant clinical potential by ensuring consistent therapeutic levels, reducing side effects, and potentially improving patient outcomes. Overall, the study highlights the effectiveness of LCD technology in printing the stent platform using PLA resin. The results demonstrated a significant cyclosporine loading with a sustained release profile without any stent coating procedure.

Authors

Institutions

Publication Details

Journal
AAPS PharmSciTech
Published
2025-03-06
DOI
https://doi.org/10.1208/s12249-025-03066-y
Primary Topic
Coronary Interventions and Diagnostics
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

In Vitro Characterization of a Cyclosporine Eluted Polymeric Stent Manufactured by Liquid Crystal Display 3D Technology

Saeideh Allahyari, Ali Nokhodchi, Saeed Mousazadeh
AAPS PharmSciTech
Coronary Interventions and Diagnostics
article

In Vitro Characterization of a Cyclosporine Eluted Polymeric Stent Manufactured by Liquid Crystal Display 3D Technology

Saeideh Allahyari, Ali Nokhodchi, Saeed Mousazadeh
article en

Abstract

Coronary artery blockage, the most common cardiovascular problem, is often treated with drug-eluting stents (DES). This study aims to address the main limitation of traditional angioplasty therapy. Thus, designing, fabricating, and analyzing a novel drug-eluting polymeric stent using liquid crystal display (LCD) technology may potentially represent an innovative approach to DES in the near future. Therefore, a poly (lactic acid) (PLA) based 3D-printed stent was designed using SolidWorks software and fabricated using the liquid crystal display (LCD) method. The cyclosporine-loaded stent was prepared and analyzed using optical microscopy, differential scanning calorimetry (DSC), and Fourier transform infrared spectroscopy (FTIR). Loading efficiency percentage and release characteristics were estimated. The polymeric stent platform was successfully designed and manufactured using PLA resin. Based on the characterization of cyclosporine eluting stent, a loading efficiency of 48.66 ± 5.92% was estimated through the immersion method. The FTIR and DSC results confirmed molecular interactions between cyclosporine and the PLA-based 3D-printed stent compared with physical mixture formulations. A sustained release profile of cyclosporine was also observed where approximately 50% of the drug was released within the first three hours. The sustained-release profile, characterized by the absence of a burst release, holds significant clinical potential by ensuring consistent therapeutic levels, reducing side effects, and potentially improving patient outcomes. Overall, the study highlights the effectiveness of LCD technology in printing the stent platform using PLA resin. The results demonstrated a significant cyclosporine loading with a sustained release profile without any stent coating procedure.

AAPS PharmSciTechVol. 26(3)
University of Sussex (GB), Ardabil University of Medical Sciences (IR)
Openalex Percentile: Top 93%
Coronary Interventions and Diagnostics
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.