Performance Evaluation of Automated Urine Analyzers: Laura XL

Urinalysis is the third most used diagnostic screening test in clinical practice. Manual urine analysis is labor-intensive and requires experienced staff for accurate results and interpretation. New-generation automated urianalysers have been introduced with microscopic analysis. The aim of this study was to evaluate the results of LauraXL, automated Urine Analyzer and validate it with UF 4000 Sysmex and manual method. Method: A cross-sectional study was conducted on 107 urine samples. Laura XL works on image-based microscopy, whereas UF 4000 works on the principle of flow cytometry. Epithelial cells (EC), red blood cells (RBC), white blood cells (WBC), crystals, cast, and yeast from both machines were compared with manual microscopy. Results: The degree of concordance was analyzed. UF4000 showed greater agreement than Laura for WBC, EC, and Crystals, kappa as 0.420, 0.238, and 0.437 respectively (p <0.0001.) Yeast showed substantial agreement with kappa 0.650(UF 4000) and 0.643 (LauraXL). The agreement of Laura with Manual was greater than UF4000 for RBC (kappa 0.360 and p<0.0001). UF4000 revealed a sensitivity of >95% for WBC, EC, CAST, Crystal, and Yeast. Similar results were observed for Laura except for Crystal (sensitivity: 81%). Conclusion: Both automatic urine analyzers exhibited comparable performances and strongly correlated with manual microscopy. While manual urinalysis remains critical for diagnosis, automated systems offer enhanced efficiency, accuracy, and reliability, making them indispensable in modern clinical laboratories.

Authors

Publication Details

Journal
South Eastern European Journal of Public Health
Published
2025-03-02
DOI
https://doi.org/10.70135/seejph.vi.5284
Primary Topic
Healthcare Technology and Patient Monitoring
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Performance Evaluation of Automated Urine Analyzers: Laura XL

Archana Buch, P Ratnakar, Poonam Lalla, Shirish Chandanwale et al.
South Eastern European Journal of Public Health
Healthcare Technology and Patient Monitoring
article

Performance Evaluation of Automated Urine Analyzers: Laura XL

Archana Buch, P Ratnakar, Poonam Lalla, Shirish Chandanwale, Charusheela R. Gore, Abhinav Shetty, Tejaswini Shrirao
article en

Abstract

Urinalysis is the third most used diagnostic screening test in clinical practice. Manual urine analysis is labor-intensive and requires experienced staff for accurate results and interpretation. New-generation automated urianalysers have been introduced with microscopic analysis. The aim of this study was to evaluate the results of LauraXL, automated Urine Analyzer and validate it with UF 4000 Sysmex and manual method. Method: A cross-sectional study was conducted on 107 urine samples. Laura XL works on image-based microscopy, whereas UF 4000 works on the principle of flow cytometry. Epithelial cells (EC), red blood cells (RBC), white blood cells (WBC), crystals, cast, and yeast from both machines were compared with manual microscopy. Results: The degree of concordance was analyzed. UF4000 showed greater agreement than Laura for WBC, EC, and Crystals, kappa as 0.420, 0.238, and 0.437 respectively (p <0.0001.) Yeast showed substantial agreement with kappa 0.650(UF 4000) and 0.643 (LauraXL). The agreement of Laura with Manual was greater than UF4000 for RBC (kappa 0.360 and p<0.0001). UF4000 revealed a sensitivity of >95% for WBC, EC, CAST, Crystal, and Yeast. Similar results were observed for Laura except for Crystal (sensitivity: 81%). Conclusion: Both automatic urine analyzers exhibited comparable performances and strongly correlated with manual microscopy. While manual urinalysis remains critical for diagnosis, automated systems offer enhanced efficiency, accuracy, and reliability, making them indispensable in modern clinical laboratories.

South Eastern European Journal of Public Health
Openalex Percentile: Top 93%
Healthcare Technology and Patient Monitoring
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.