Coronary Plaque Quantification with Ultrahigh-Spatial-Resolution Photon-counting Detector CT: Intraindividual Comparison with Energy-integrating Detector CT

Background Other than enhancing the accuracy of stenosis measurements, the improved spatial resolution of photon-counting detector (PCD) CT may have an impact on quantitative plaque assessment at coronary CT angiography (CCTA). Purpose To evaluate the effect of PCD CT on coronary plaque quantification and characterization compared with that of energy-integrating detector (EID) CT. Materials and Methods Consecutive participants undergoing clinically indicated CCTA at EID CT (192 × 0.6-mm collimation) were enrolled to undergo ultrahigh-spatial-resolution (UHR) PCD CT (120 × 0.2-mm collimation) within 30 days. PCD CT was performed using equivalent or lower CT dose index and equivalent contrast media volume as the clinical scan. Total, calcified, fibrotic, and low-attenuation coronary plaque volumes were quantified and compared between scanners. Intra- and interreader reproducibility was assessed for both systems. Results A total of 164 plaques from 48 participants were segmented on both scans. Total plaque volume was lower at PCD CT compared with EID CT (723.5 mm<sup>3</sup> [IQR, 500.6-1184.7 mm<sup>3</sup>] vs 1084.7 mm<sup>3</sup> [IQR, 710.7-1609.8 mm<sup>3</sup>]; <i>P</i> < .001). UHR-based segmentations produced lower fibrotic (325.4 mm<sup>3</sup> [IQR, 151.7-519.2 mm<sup>3</sup>] vs 627.7 mm<sup>3</sup> [IQR, 385.8-795.1 mm<sup>3</sup>], respectively; <i>P</i> < .001) and higher low-attenuation plaque volumes (72.1 mm<sup>3</sup> [IQR, 38.6-161.9 mm<sup>3</sup>] vs 58.1 mm<sup>3</sup> [IQR, 23.4-102.3 mm<sup>3</sup>], respectively; <i>P</i> = .004) than EID CT-based measurements. Calcified plaque volumes did not differ significantly between PCD CT and EID CT (344.5 mm<sup>3</sup> [IQR, 174.3-605.7 mm<sup>3</sup>] vs 342.1 mm<sup>3</sup> [IQR, 180.4-607.5 mm<sup>3</sup>], respectively; <i>P</i> = .13). Total, calcified, and fibrotic plaque volumes demonstrated excellent agreement between repeated measurements and between readers for both PCD CT and EID CT (all intraclass correlation coefficients [ICCs] > 0.90). Whereas low-attenuation plaque volume had strong intrareader (ICC, 0.84; 95% CI: 0.57, 0.94) and interreader (ICC, 0.92; 95% CI: 0.81, 0.97) agreements for PCD CT, EID CT showed only moderate (ICC, 0.62; 95% CI: 0.11, 0.86) and poor (ICC, 0.47; 95% CI: 0.01, 0.79) intrareader and interreader reproducibility. Conclusion Compared with EID CT, PCD CT UHR imaging reduced segmented coronary plaque volume by nearly one-third and improved reproducibility of low-attenuation plaque measurements. © RSNA, 2025 <i>Supplemental material is available for this article.</i>

Authors

Institutions

Publication Details

Journal
Radiology
Published
2025-03-01
DOI
https://doi.org/10.1148/radiol.241479
Citations
23
Primary Topic
Advanced X-ray and CT Imaging
Type
article
Field-Weighted Citation Impact
15.50
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Coronary Plaque Quantification with Ultrahigh-Spatial-Resolution Photon-counting Detector CT: Intraindividual Comparison with Energy-integrating Detector CT

Nicola Fink, Emese Zsarnóczay, Ákos Varga‐Szemes, Chiara Gnasso et al.
23 citations
Radiology
Advanced X-ray and CT Imaging
15.50
article

Coronary Plaque Quantification with Ultrahigh-Spatial-Resolution Photon-counting Detector CT: Intraindividual Comparison with Energy-integrating Detector CT

Nicola Fink, Emese Zsarnóczay, Ákos Varga‐Szemes, Chiara Gnasso, Ismail Kabakus, Dmitrij Kravchenko, Pál Surányi, Moritz C. Halfmann, U. Joseph Schoepf, Bálint Szilveszter, Giuseppe Tremamunno, Milán Vecsey-Nagy, Tilman Emrich, Jim O’Doherty, Pál Maurovich‐Horvat
article en
23 citations

Abstract

Background Other than enhancing the accuracy of stenosis measurements, the improved spatial resolution of photon-counting detector (PCD) CT may have an impact on quantitative plaque assessment at coronary CT angiography (CCTA). Purpose To evaluate the effect of PCD CT on coronary plaque quantification and characterization compared with that of energy-integrating detector (EID) CT. Materials and Methods Consecutive participants undergoing clinically indicated CCTA at EID CT (192 × 0.6-mm collimation) were enrolled to undergo ultrahigh-spatial-resolution (UHR) PCD CT (120 × 0.2-mm collimation) within 30 days. PCD CT was performed using equivalent or lower CT dose index and equivalent contrast media volume as the clinical scan. Total, calcified, fibrotic, and low-attenuation coronary plaque volumes were quantified and compared between scanners. Intra- and interreader reproducibility was assessed for both systems. Results A total of 164 plaques from 48 participants were segmented on both scans. Total plaque volume was lower at PCD CT compared with EID CT (723.5 mm<sup>3</sup> [IQR, 500.6-1184.7 mm<sup>3</sup>] vs 1084.7 mm<sup>3</sup> [IQR, 710.7-1609.8 mm<sup>3</sup>]; <i>P</i> < .001). UHR-based segmentations produced lower fibrotic (325.4 mm<sup>3</sup> [IQR, 151.7-519.2 mm<sup>3</sup>] vs 627.7 mm<sup>3</sup> [IQR, 385.8-795.1 mm<sup>3</sup>], respectively; <i>P</i> < .001) and higher low-attenuation plaque volumes (72.1 mm<sup>3</sup> [IQR, 38.6-161.9 mm<sup>3</sup>] vs 58.1 mm<sup>3</sup> [IQR, 23.4-102.3 mm<sup>3</sup>], respectively; <i>P</i> = .004) than EID CT-based measurements. Calcified plaque volumes did not differ significantly between PCD CT and EID CT (344.5 mm<sup>3</sup> [IQR, 174.3-605.7 mm<sup>3</sup>] vs 342.1 mm<sup>3</sup> [IQR, 180.4-607.5 mm<sup>3</sup>], respectively; <i>P</i> = .13). Total, calcified, and fibrotic plaque volumes demonstrated excellent agreement between repeated measurements and between readers for both PCD CT and EID CT (all intraclass correlation coefficients [ICCs] > 0.90). Whereas low-attenuation plaque volume had strong intrareader (ICC, 0.84; 95% CI: 0.57, 0.94) and interreader (ICC, 0.92; 95% CI: 0.81, 0.97) agreements for PCD CT, EID CT showed only moderate (ICC, 0.62; 95% CI: 0.11, 0.86) and poor (ICC, 0.47; 95% CI: 0.01, 0.79) intrareader and interreader reproducibility. Conclusion Compared with EID CT, PCD CT UHR imaging reduced segmented coronary plaque volume by nearly one-third and improved reproducibility of low-attenuation plaque measurements. © RSNA, 2025 <i>Supplemental material is available for this article.</i>

RadiologyVol. 314(3)
Semmelweis University (HU), Siemens (Germany) (DE), Medical University of South Carolina (US), Vita-Salute San Raffaele University (IT), Johannes Gutenberg University Mainz (DE), University Hospital Bonn (DE), LMU Klinikum (DE), University Medical Center of the Johannes Gutenberg University Mainz (DE), Medical Solutions (GB), Istituti di Ricovero e Cura a Carattere Scientifico (IT), Istituto di Ricovero e Cura a Carattere Scientifico San Raffaele (IT), Ludwig-Maximilians-Universität München (DE), Sapienza University of Rome (IT)
Affordable and clean energy
Openalex Percentile: Top 1%
Advanced X-ray and CT Imaging
15.50
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.