Precise and Accurate Assessment of the Copper Scrap Composition by the X-Ray Fluorescence Spectrometry

This paper evaluates the possibility of analyzing the composition of high-quality copper scrap with X-ray fluorescence spectrometry (XRF) instead of electrogravimetry combined with flame atomic absorption spectroscopy (EG+FAAS) method. The evaluation of the methods was performed on three real copper scraps, the composition of which were estimated in average samples taken after re-melting. Traceability of XRF results was ensured by the use of new certified reference materials (CRMs) dedicated to copper scrap analysis. The copper content results obtained by XRF were characterized by high agreement with the reference results obtained by EG+FAAS method. The estimated expanded uncertainty for Cu of both methods was 0.23%. In addition to significant time savings without compromising accuracy, the XRF method additionally provided information on the content of 12 other elements, such as Sn, Zn, Co, Cd, Sb, Ni, Fe, Pb, Bi, Ag, Al, and P. This may make the XRF method attractive compared to the commonly used EG+FAAS method.

Authors

Institutions

Publication Details

Journal
Recycling
Published
2025-03-04
DOI
https://doi.org/10.3390/recycling10020036
Citations
1
Primary Topic
Electron and X-Ray Spectroscopy Techniques
Type
article
Field-Weighted Citation Impact
0.66

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Precise and Accurate Assessment of the Copper Scrap Composition by the X-Ray Fluorescence Spectrometry

Tadeusz Gorewoda, Jacek Anyszkiewicz, Sonia Kasierot, Justyna Kostrzewa
1 citations
Recycling
Electron and X-Ray Spectroscopy Techniques
0.66
article

Precise and Accurate Assessment of the Copper Scrap Composition by the X-Ray Fluorescence Spectrometry

Tadeusz Gorewoda, Jacek Anyszkiewicz, Sonia Kasierot, Justyna Kostrzewa
article en
1 citations

Abstract

This paper evaluates the possibility of analyzing the composition of high-quality copper scrap with X-ray fluorescence spectrometry (XRF) instead of electrogravimetry combined with flame atomic absorption spectroscopy (EG+FAAS) method. The evaluation of the methods was performed on three real copper scraps, the composition of which were estimated in average samples taken after re-melting. Traceability of XRF results was ensured by the use of new certified reference materials (CRMs) dedicated to copper scrap analysis. The copper content results obtained by XRF were characterized by high agreement with the reference results obtained by EG+FAAS method. The estimated expanded uncertainty for Cu of both methods was 0.23%. In addition to significant time savings without compromising accuracy, the XRF method additionally provided information on the content of 12 other elements, such as Sn, Zn, Co, Cd, Sb, Ni, Fe, Pb, Bi, Ag, Al, and P. This may make the XRF method attractive compared to the commonly used EG+FAAS method.

RecyclingVol. 10(2)
Łukasiewicz Research Network - Institute of Non-Ferrous Metals (PL), Łukasiewicz Research Network (PL)
Narodowe Centrum Badań i Rozwoju
Openalex Percentile: Top 35%
Electron and X-Ray Spectroscopy Techniques
0.66
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.