Wearable Device for Blood Pressure Monitoring Based on a Novel Algorithm With a Few Morphological Feature Parameters of Pulse Waves

We proposed an algorithm for blood pressure estimation based on a few morphological feature parameters of pulse waves to solve the problem of over-reliance on high-quality pulse waves (involving numerous pulse wave morphological feature parameters) of blood pressure estimation based on single-channel photoplethysmography (PPG). We successfully implemented the algorithm in a self-developed miniaturized wearable device. Statistical analysis revealed that in the algorithm, a parameter calculated from pulse pressure difference (DP) was highly correlated (<inline-formula xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink"> <tex-math notation="LaTeX">${R}^{{2}} = {0.942}$ </tex-math></inline-formula>) with systolic blood pressure (Ps). We defined this parameter as k. We used multiple linear regression to establish a mapping relationship between the morphological feature parameters of pulse waves and DP. This relationship had a correlation coefficient of <inline-formula xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink"> <tex-math notation="LaTeX">${R}^{{2}} = {0.667}$ </tex-math></inline-formula>, <inline-formula xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink"> <tex-math notation="LaTeX">${P}~\lt 0.001$ </tex-math></inline-formula>. We built a miniaturized platform for blood pressure measurement and developed a reflective PPG sensor packaged with polydimethylsiloxane to enable comfortable wearable blood pressure measurement. In uncalibrated randomized experiments, we first evaluated the performance of the algorithm on separate datasets. A total of 116 objects showed a mean absolute error (MAE) of 11.15 and 8.29 mmHg for Ps and diastolic blood pressure (Pd), respectively. In a human experiment with 31 subjects, the errors for Ps and Pd were <inline-formula xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink"> <tex-math notation="LaTeX">$5.29~\pm ~10.92$ </tex-math></inline-formula> and <inline-formula xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink"> <tex-math notation="LaTeX">$- 2.77~\pm ~8.71$ </tex-math></inline-formula> mmHg, respectively. The results of multistate continuous blood pressure monitoring experiments improved after calibration was performed once. In a prolonged blood pressure monitoring experiment, the errors for the Ps and Pd of subjects were <inline-formula xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink"> <tex-math notation="LaTeX">$1.11~\pm ~7.05$ </tex-math></inline-formula> and <inline-formula xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink"> <tex-math notation="LaTeX">$0.19~\pm ~2.69$ </tex-math></inline-formula> mmHg, respectively. The results of both experiments conducted in this work were close to the international standard (<inline-formula xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink"> <tex-math notation="LaTeX">$5~\pm ~8$ </tex-math></inline-formula> mmHg) issued by the American Association for the Advancement of Medical Instrumentation. Our algorithm also showed promising results in continuous blood pressure monitoring experiments after exercise and under breath-holding conditions.

Authors

Institutions

Publication Details

Journal
IEEE Sensors Journal
Published
2025-03-04
DOI
https://doi.org/10.1109/jsen.2025.3545423
Citations
3
Primary Topic
Non-Invasive Vital Sign Monitoring
Type
article
Field-Weighted Citation Impact
2.02

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Wearable Device for Blood Pressure Monitoring Based on a Novel Algorithm With a Few Morphological Feature Parameters of Pulse Waves

Di Wu, Zhengchun Liu, Long Mo, Yu Peng et al.
3 citations
IEEE Sensors Journal
Non-Invasive Vital Sign Monitoring
2.02
article

Wearable Device for Blood Pressure Monitoring Based on a Novel Algorithm With a Few Morphological Feature Parameters of Pulse Waves

Di Wu, Zhengchun Liu, Long Mo, Yu Peng, Yanke Zhang, Junwei Zhang, Chengqi Zhang, Chang Zeng, Weixuan Shao, Guannan Peng
article en
3 citations

Abstract

We proposed an algorithm for blood pressure estimation based on a few morphological feature parameters of pulse waves to solve the problem of over-reliance on high-quality pulse waves (involving numerous pulse wave morphological feature parameters) of blood pressure estimation based on single-channel photoplethysmography (PPG). We successfully implemented the algorithm in a self-developed miniaturized wearable device. Statistical analysis revealed that in the algorithm, a parameter calculated from pulse pressure difference (DP) was highly correlated (<inline-formula xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink"> <tex-math notation="LaTeX">${R}^{{2}} = {0.942}$ </tex-math></inline-formula>) with systolic blood pressure (Ps). We defined this parameter as k. We used multiple linear regression to establish a mapping relationship between the morphological feature parameters of pulse waves and DP. This relationship had a correlation coefficient of <inline-formula xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink"> <tex-math notation="LaTeX">${R}^{{2}} = {0.667}$ </tex-math></inline-formula>, <inline-formula xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink"> <tex-math notation="LaTeX">${P}~\lt 0.001$ </tex-math></inline-formula>. We built a miniaturized platform for blood pressure measurement and developed a reflective PPG sensor packaged with polydimethylsiloxane to enable comfortable wearable blood pressure measurement. In uncalibrated randomized experiments, we first evaluated the performance of the algorithm on separate datasets. A total of 116 objects showed a mean absolute error (MAE) of 11.15 and 8.29 mmHg for Ps and diastolic blood pressure (Pd), respectively. In a human experiment with 31 subjects, the errors for Ps and Pd were <inline-formula xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink"> <tex-math notation="LaTeX">$5.29~\pm ~10.92$ </tex-math></inline-formula> and <inline-formula xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink"> <tex-math notation="LaTeX">$- 2.77~\pm ~8.71$ </tex-math></inline-formula> mmHg, respectively. The results of multistate continuous blood pressure monitoring experiments improved after calibration was performed once. In a prolonged blood pressure monitoring experiment, the errors for the Ps and Pd of subjects were <inline-formula xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink"> <tex-math notation="LaTeX">$1.11~\pm ~7.05$ </tex-math></inline-formula> and <inline-formula xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink"> <tex-math notation="LaTeX">$0.19~\pm ~2.69$ </tex-math></inline-formula> mmHg, respectively. The results of both experiments conducted in this work were close to the international standard (<inline-formula xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink"> <tex-math notation="LaTeX">$5~\pm ~8$ </tex-math></inline-formula> mmHg) issued by the American Association for the Advancement of Medical Instrumentation. Our algorithm also showed promising results in continuous blood pressure monitoring experiments after exercise and under breath-holding conditions.

IEEE Sensors JournalVol. 25(8)
Central South University (CN), Xiangya Hospital Central South University (CN)
National Natural Science Foundation of China
Openalex Percentile: Top 15%
Non-Invasive Vital Sign Monitoring
2.02
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.