Mycoplasma pneumoniae (MP) is the main culprit of community-acquired pneumonia. Commonly used laboratory testing methods have many shortcomings. Serological diagnosis has low sensitivity, causing false negatives, while a quantitative real-time polymerase chain reaction (qPCR) requires large equipment and professional staff. To make up for these shortcomings, we proposed a label-free, low-cost, and small-sized ion-sensitive field-effect transistor (ISFET) array based on a low-buffered loop-mediated isothermal amplification (LAMP) assay. A complementary metal oxide semiconductor (CMOS)-based ISFET array with 512 × 512 sensors was used in this system, which responds specifically to H<sup>+</sup> with a sensitivity of 365.7 mV/pH. For on-chip amplification, a low-buffered LAMP system designed for the conserved sequences of two genes, <i>CARDS</i> and <i>gyrB</i>, was applied. The rapid release of large amounts of H<sup>+</sup> in the low-buffered LAMP solution led to a speedy increase in electrical signals captured by the ISFET array, eliminating the need for a sophisticated temperature cycling and optical system. The on-chip results showed that the device can accurately complete MP detection with a detection limit of about 10<sup>3</sup> copies/mL (approximately 1 copy per reaction). In the final clinical validation, the detection results of eight throat swab samples using the ISFET sensors were fully consistent with the clinical laboratory diagnostic outcomes, confirming the accuracy and reliability of the ISFET sensors for use in clinical settings. And the entire process from sample lysis to result interpretation takes about 60 min. This platform has potential to be used for the point-of-care testing (POCT) of pathogen infections, providing a basis for the timely adjustment of diagnosis and treatment plans.

Authors

Institutions

Publication Details

Journal
Sensors
Published
2025-03-04
DOI
https://doi.org/10.3390/s25051562
Citations
3
Primary Topic
Biosensors and Analytical Detection
Type
article
Field-Weighted Citation Impact
2.02

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

ISFET Biosensor with Loop-Mediated Isothermal Amplification for Electronic Rapid Detection of Mycoplasma Pneumoniae

Walter Hu, Hao Bai, Jie Zou, Weiyi Bai et al.
3 citations
Sensors
Biosensors and Analytical Detection
2.02
article

ISFET Biosensor with Loop-Mediated Isothermal Amplification for Electronic Rapid Detection of Mycoplasma Pneumoniae

Walter Hu, Hao Bai, Jie Zou, Weiyi Bai, Limei Zhang, Lei Wang, Zhifeng Zhang, Jianlong Li, Yan Lin, Jie Hu, Yan Shen
article en
3 citations

Abstract

Mycoplasma pneumoniae (MP) is the main culprit of community-acquired pneumonia. Commonly used laboratory testing methods have many shortcomings. Serological diagnosis has low sensitivity, causing false negatives, while a quantitative real-time polymerase chain reaction (qPCR) requires large equipment and professional staff. To make up for these shortcomings, we proposed a label-free, low-cost, and small-sized ion-sensitive field-effect transistor (ISFET) array based on a low-buffered loop-mediated isothermal amplification (LAMP) assay. A complementary metal oxide semiconductor (CMOS)-based ISFET array with 512 × 512 sensors was used in this system, which responds specifically to H<sup>+</sup> with a sensitivity of 365.7 mV/pH. For on-chip amplification, a low-buffered LAMP system designed for the conserved sequences of two genes, <i>CARDS</i> and <i>gyrB</i>, was applied. The rapid release of large amounts of H<sup>+</sup> in the low-buffered LAMP solution led to a speedy increase in electrical signals captured by the ISFET array, eliminating the need for a sophisticated temperature cycling and optical system. The on-chip results showed that the device can accurately complete MP detection with a detection limit of about 10<sup>3</sup> copies/mL (approximately 1 copy per reaction). In the final clinical validation, the detection results of eight throat swab samples using the ISFET sensors were fully consistent with the clinical laboratory diagnostic outcomes, confirming the accuracy and reliability of the ISFET sensors for use in clinical settings. And the entire process from sample lysis to result interpretation takes about 60 min. This platform has potential to be used for the point-of-care testing (POCT) of pathogen infections, providing a basis for the timely adjustment of diagnosis and treatment plans.

SensorsVol. 25(5)
Sichuan University (CN), West China Hospital of Sichuan University (CN)
National Natural Science Foundation of China, China Postdoctoral Science Foundation
Good health and well-being
Openalex Percentile: Top 15%
Biosensors and Analytical Detection
2.02
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.