Aims: This in silico study aimed to investigate the effect of implant–abutment contact surfaces on the stress generation of Morse taper implants under oblique loading. Materials and methods: Three-dimensional finite element models of Bone-Level and Tissue-Level implants were simulated with Standard and Partial contacts between the abutment and implant. The dimensional parameters followed the ISO 14801 guidelines, and an oblique load of 300 N was applied to the implants. The von Mises stress was acquired. Results: The Tissue-Level design showed a significant difference in the stress level when the connection with the implant, abutment, and screw was Partial. For the implant fixture, abutment, and screw, the Tissue-Level design showed 13% more stress in the implant, abutment, and screw when the connection was Partial. The Bone-Level design did not affect the connection and showed an overall 42% lower stress than the Tissue-Level design for the implant fixture. However, in the screw, there was a difference between the Bone-Level implants with a Standard and Partial connection. In contrast, for the Tissue-Level implant, this difference was less evident with higher stress peaks in the entire set. Conclusion: To achieve optimal outcomes, it is highly recommended to use original abutments, as they provide a more precise fit. The stress peaks were notably lower in Bone-Level implants compared to Tissue-Level implants. Furthermore, an implant–abutment connection with more contacting areas significantly reduced stress concentration, especially in Tissue-Level implant designs. By choosing well-fitting abutments, one can ensure more stable and durable implant performance with less stress.

Authors

Institutions

Publication Details

Journal
Applied Sciences
Published
2025-03-04
DOI
https://doi.org/10.3390/app15052744
Citations
4
Primary Topic
Dental Implant Techniques and Outcomes
Type
article
Field-Weighted Citation Impact
7.16
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

The Effect of Implant–Abutment Contact Area on the Stress Generation of Bone-Level and Tissue-Level Implants

Cornelis J. Kleverlaan, Amanda Maria de Oliveira Dal Piva, Albert J. Feilzer, Erik J. Blom et al.
4 citations
Applied Sciences
Dental Implant Techniques and Outcomes
7.16
article

The Effect of Implant–Abutment Contact Area on the Stress Generation of Bone-Level and Tissue-Level Implants

Cornelis J. Kleverlaan, Amanda Maria de Oliveira Dal Piva, Albert J. Feilzer, Erik J. Blom, Pietro Ausiello, Maria Bruhnke, João Paulo Mendes Tribst, Nilüfer Özkara
article en
4 citations

Abstract

Aims: This in silico study aimed to investigate the effect of implant–abutment contact surfaces on the stress generation of Morse taper implants under oblique loading. Materials and methods: Three-dimensional finite element models of Bone-Level and Tissue-Level implants were simulated with Standard and Partial contacts between the abutment and implant. The dimensional parameters followed the ISO 14801 guidelines, and an oblique load of 300 N was applied to the implants. The von Mises stress was acquired. Results: The Tissue-Level design showed a significant difference in the stress level when the connection with the implant, abutment, and screw was Partial. For the implant fixture, abutment, and screw, the Tissue-Level design showed 13% more stress in the implant, abutment, and screw when the connection was Partial. The Bone-Level design did not affect the connection and showed an overall 42% lower stress than the Tissue-Level design for the implant fixture. However, in the screw, there was a difference between the Bone-Level implants with a Standard and Partial connection. In contrast, for the Tissue-Level implant, this difference was less evident with higher stress peaks in the entire set. Conclusion: To achieve optimal outcomes, it is highly recommended to use original abutments, as they provide a more precise fit. The stress peaks were notably lower in Bone-Level implants compared to Tissue-Level implants. Furthermore, an implant–abutment connection with more contacting areas significantly reduced stress concentration, especially in Tissue-Level implant designs. By choosing well-fitting abutments, one can ensure more stable and durable implant performance with less stress.

Applied SciencesVol. 15(5)
Academic Center for Dentistry Amsterdam (NL), Humboldt-Universität zu Berlin (DE), University of Naples Federico II (IT), Freie Universität Berlin (DE), Charité - Universitätsmedizin Berlin (DE), University of Amsterdam (NL)
Openalex Percentile: Top 4%
Dental Implant Techniques and Outcomes
7.16
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.