A Novel Trans-Impedance Matrix (TIM) Abnormality Pattern in Cochlear Implants

In our clinical setting, we have identified a novel pattern of Trans-Impedance Matrix (TIM) measurement that we call 'scatter', a measure characterised by a loss of definition in the heat and line maps. <b>Objective:</b> the aim of this study was to describe the basic characteristics of the anomaly pattern. The secondary purpose is to evaluate correlations between the "scatter" pattern and normal TIM by considering different parameters. <b>Methods:</b> the experimental sample, therefore, consisted of 565 patients (81.1% of people with a checked TIM at follow-up; M: 279, F: 286 and mean age: 27 years (sd 26). The scatter pattern was found in 55 devices (9.7%). We classified this pattern as severe (20 devices) or mild (35 devices) according to the visual extent of the abnormality. We considered the visual extension of the pattern, device lifetime, type of internal part of the CI, and auditory performance (speech audiometry in quiet at 65 dB and in noise-Ita Matrix Sentence Test). We also analysed two quantitive parameters: Shannon entropy and exponential decay. <b>Results:</b> a difference was found in these two quantitative parameters between the severe scatter, mild scatter, and normal TIM groups (<i>p</i>-value < 0.0001). The severe scatter group seems to be related to the type of device (CI24RE and CI512) and long device life (average 133 months); it did not show differences in audiology performances compared to the other groups. <b>Conclusions:</b> this result gives a numerical validation to the more subjective visual inspection approach. The scatter pattern is a novel, previously undescribed abnormality of TIM. It can vary from moderate to severe. A numerical basis to validate the inspection approach is described here.

Authors

Institutions

Publication Details

Journal
Audiology Research
Published
2025-03-02
DOI
https://doi.org/10.3390/audiolres15020024
Citations
1
Primary Topic
Hearing Loss and Rehabilitation
Type
article
Field-Weighted Citation Impact
1.43
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

A Novel Trans-Impedance Matrix (TIM) Abnormality Pattern in Cochlear Implants

Patrizia Frontera, Domenico Cuda, Sara Ghiselli, Erica Pizzol et al.
1 citations
Audiology Research
Hearing Loss and Rehabilitation
1.43
article

A Novel Trans-Impedance Matrix (TIM) Abnormality Pattern in Cochlear Implants

Patrizia Frontera, Domenico Cuda, Sara Ghiselli, Erica Pizzol, Daria Salsi
article en
1 citations

Abstract

In our clinical setting, we have identified a novel pattern of Trans-Impedance Matrix (TIM) measurement that we call 'scatter', a measure characterised by a loss of definition in the heat and line maps. <b>Objective:</b> the aim of this study was to describe the basic characteristics of the anomaly pattern. The secondary purpose is to evaluate correlations between the "scatter" pattern and normal TIM by considering different parameters. <b>Methods:</b> the experimental sample, therefore, consisted of 565 patients (81.1% of people with a checked TIM at follow-up; M: 279, F: 286 and mean age: 27 years (sd 26). The scatter pattern was found in 55 devices (9.7%). We classified this pattern as severe (20 devices) or mild (35 devices) according to the visual extent of the abnormality. We considered the visual extension of the pattern, device lifetime, type of internal part of the CI, and auditory performance (speech audiometry in quiet at 65 dB and in noise-Ita Matrix Sentence Test). We also analysed two quantitive parameters: Shannon entropy and exponential decay. <b>Results:</b> a difference was found in these two quantitative parameters between the severe scatter, mild scatter, and normal TIM groups (<i>p</i>-value < 0.0001). The severe scatter group seems to be related to the type of device (CI24RE and CI512) and long device life (average 133 months); it did not show differences in audiology performances compared to the other groups. <b>Conclusions:</b> this result gives a numerical validation to the more subjective visual inspection approach. The scatter pattern is a novel, previously undescribed abnormality of TIM. It can vary from moderate to severe. A numerical basis to validate the inspection approach is described here.

Audiology ResearchVol. 15(2)
University of Parma (IT), Azienda Unità Sanitaria Locale Piacenza (IT)
Openalex Percentile: Top 22%
Hearing Loss and Rehabilitation
1.43
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

A Novel Trans-Impedance Matrix (TIM) Abnormality Pattern in Cochlear Implants — Patrizia Frontera, Domenico Cuda, et al. · Audiology Research (2025) | TGRS Research Map | TGRS