A multicenter study comparing a point-of-care blood glucose metre with a blood gas analyser in infants by the Shikoku Neonatal Medical Research Group

BackgroundThere have been no reports on blood glucose metres for measurements in a wide variety of infant patients, from very preterm infants to mature infants and from the early neonatal period onwards. In this study, we evaluated the accuracy of the Glutest Mint II, a point-of-care blood glucose metre, by comparing blood glucose levels measured by this device with those measured by a blood gas analyser in infants of all gestational ages and birth weights at a number of time points after birth.MethodsInfants aged 22 weeks and 0 days of gestation or older who were born at any of six tertiary neonatal intensive care units between March 2022 and January 2023 and needed blood glucose monitoring were enrolled. Samples were collected into capillary tubes when the physician determined that a blood glucose test was necessary and could be taken at any time after birth and any number of times from the same individual. Blood glucose was measured using a Glutest Mint II and then using a blood gas analyser.ResultsIn total, 2943 blood glucose points were measured in 285 infants. Blood glucose levels measured using the Glutest Mint II were significantly correlated with those measured using a blood gas analyser. Neonatal-specific parameters such as hematocrit and total serum bilirubin levels may not have an effect.ConclusionsThe Glutest Mint II device can measure blood glucose levels with very high accuracy in the range used in the neonatal setting, comparable to the blood gas analyser.

Authors

Institutions

Publication Details

Journal
Annals of Clinical Biochemistry International Journal of Laboratory Medicine
Published
2025-03-04
DOI
https://doi.org/10.1177/00045632251326460
Citations
1
Primary Topic
Hyperglycemia and glycemic control in critically ill and hospitalized patients
Type
article
Field-Weighted Citation Impact
1.29
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

A multicenter study comparing a point-of-care blood glucose metre with a blood gas analyser in infants by the Shikoku Neonatal Medical Research Group

Kenji Miyamoto, Masaaki Ota, Yusei Nakata, Takaaki Sadamura et al.
1 citations
Annals of Clinical Biochemistry International Journal of Laboratory Medicine
Hyperglycemia and glycemic control in critically ill and hospitalized patients
1.29
article

A multicenter study comparing a point-of-care blood glucose metre with a blood gas analyser in infants by the Shikoku Neonatal Medical Research Group

Kenji Miyamoto, Masaaki Ota, Yusei Nakata, Takaaki Sadamura, Noriko N. Miura, Shinji Nakamura, Kenichi Suga, Natsumi Okamoto, Toru Kuboi, Yoko Tadatomo, Takashi Kusaka, Kosuke Koyano, Akiko Nakano, Nana Kawaguchi, Kosei Takada, Masashiro Sugino
article en
1 citations

Abstract

BackgroundThere have been no reports on blood glucose metres for measurements in a wide variety of infant patients, from very preterm infants to mature infants and from the early neonatal period onwards. In this study, we evaluated the accuracy of the Glutest Mint II, a point-of-care blood glucose metre, by comparing blood glucose levels measured by this device with those measured by a blood gas analyser in infants of all gestational ages and birth weights at a number of time points after birth.MethodsInfants aged 22 weeks and 0 days of gestation or older who were born at any of six tertiary neonatal intensive care units between March 2022 and January 2023 and needed blood glucose monitoring were enrolled. Samples were collected into capillary tubes when the physician determined that a blood glucose test was necessary and could be taken at any time after birth and any number of times from the same individual. Blood glucose was measured using a Glutest Mint II and then using a blood gas analyser.ResultsIn total, 2943 blood glucose points were measured in 285 infants. Blood glucose levels measured using the Glutest Mint II were significantly correlated with those measured using a blood gas analyser. Neonatal-specific parameters such as hematocrit and total serum bilirubin levels may not have an effect.ConclusionsThe Glutest Mint II device can measure blood glucose levels with very high accuracy in the range used in the neonatal setting, comparable to the blood gas analyser.

Annals of Clinical Biochemistry International Journal of Laboratory MedicineVol. 62(5)
Kagawa University (JP), CHI Health (US), Faculty (United Kingdom) (GB), Tokushima University Hospital (JP), National Hospital Organization (JP), Shikoku Medical Center for Children and Adults (JP), Ehime University (JP), Kōchi University (JP)
Good health and well-being
Openalex Percentile: Top 22%
Hyperglycemia and glycemic control in critically ill and hospitalized patients
1.29
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.