Understanding Platelet Activation in the Aeson Bioprosthetic Total Artificial Heart: Insights From Aspirin Treatment and Outcomes

This study aimed to assess platelet activation following implantation of the Aeson bioprosthetic total artificial heart (A-TAH). We monitored plasma levels of platelet activation markers in patients receiving A-TAH support (n = 16) throughout the follow-up period. Before implantation, soluble CD40 ligand (sCD40L) levels averaged 3,909.06 pg/ml (standard deviation [SD] = 3,772.37), remaining stable postimplantation at 3,964.56 pg/ml (SD = 2,198.85) during months 1-3 and at 3,519.27 pg/ml (SD = 1,647.04) during months 3-6. Similarly, P-selectin (sP-sel) levels were 35,235.36 pg/ml (SD = 14,940.47) before implantation, stabilizing to 33,158.96 pg/ml (SD = 9,023.11) (1-3 months) and 31,022.58 pg/ml (SD = 9,249.95) (3-6 months). Preimplantation platelet factor 4 (PF4) measured 2,593.47 ng/ml (SD = 2,167.85), remaining consistent at 2,136.10 ng/ml (SD = 1,264.47) (1-3 months) and 1,991.26 ng/ml (SD = 1,234.16) (3-6 months). Levels of neutrophil-activating peptide 2 (NAP2) were also steady, measuring 785.63 ng/ml (SD = 605.26) preimplantation, 935.10 ng/ml (SD = 517.73) at 1-3 months, and 907.21 ng/ml (SD = 501.96) at 3-6 months postimplantation. Importantly, neither aspirin nor heparin treatment affected these platelet biomarker levels. No correlation was observed between platelet activation marker levels and clinical outcomes such as pericardial effusion, nor with the timing of aspirin initiation and drain removal. Our findings confirm that A-TAH does not trigger platelet activation. The lack of association between aspirin, platelet activation, and clinical outcomes suggests the possibility of discontinuing antiplatelet therapy following A-TAH implantation in the future.

Authors

Institutions

Publication Details

Journal
ASAIO Journal
Published
2025-02-28
DOI
https://doi.org/10.1097/mat.0000000000002403
Citations
1
Primary Topic
Mechanical Circulatory Support Devices
Type
article
Field-Weighted Citation Impact
0.67
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Understanding Platelet Activation in the Aeson Bioprosthetic Total Artificial Heart: Insights From Aspirin Treatment and Outcomes

David Lebeaux, Erwan Flécher, Finn Gustafsson, David M. Smadja et al.
1 citations
ASAIO Journal
Mechanical Circulatory Support Devices
0.67
article

Understanding Platelet Activation in the Aeson Bioprosthetic Total Artificial Heart: Insights From Aspirin Treatment and Outcomes

David Lebeaux, Erwan Flécher, Finn Gustafsson, David M. Smadja, Marie Courbebaisse, Béatrice Parfait, Faiz Ramjankhan, Christian Latrémouille, Gérard Friedlander, Jean‐Christian Roussel, André Vincentelli, Ivan Netuka, Pascale Gaussem, Yuri Pya, Aurélien Philippe, Léa Jilet, Christophe Peronino, Joseph Roux de Bezieux, Piet Jansen
article en
1 citations

Abstract

This study aimed to assess platelet activation following implantation of the Aeson bioprosthetic total artificial heart (A-TAH). We monitored plasma levels of platelet activation markers in patients receiving A-TAH support (n = 16) throughout the follow-up period. Before implantation, soluble CD40 ligand (sCD40L) levels averaged 3,909.06 pg/ml (standard deviation [SD] = 3,772.37), remaining stable postimplantation at 3,964.56 pg/ml (SD = 2,198.85) during months 1-3 and at 3,519.27 pg/ml (SD = 1,647.04) during months 3-6. Similarly, P-selectin (sP-sel) levels were 35,235.36 pg/ml (SD = 14,940.47) before implantation, stabilizing to 33,158.96 pg/ml (SD = 9,023.11) (1-3 months) and 31,022.58 pg/ml (SD = 9,249.95) (3-6 months). Preimplantation platelet factor 4 (PF4) measured 2,593.47 ng/ml (SD = 2,167.85), remaining consistent at 2,136.10 ng/ml (SD = 1,264.47) (1-3 months) and 1,991.26 ng/ml (SD = 1,234.16) (3-6 months). Levels of neutrophil-activating peptide 2 (NAP2) were also steady, measuring 785.63 ng/ml (SD = 605.26) preimplantation, 935.10 ng/ml (SD = 517.73) at 1-3 months, and 907.21 ng/ml (SD = 501.96) at 3-6 months postimplantation. Importantly, neither aspirin nor heparin treatment affected these platelet biomarker levels. No correlation was observed between platelet activation marker levels and clinical outcomes such as pericardial effusion, nor with the timing of aspirin initiation and drain removal. Our findings confirm that A-TAH does not trigger platelet activation. The lack of association between aspirin, platelet activation, and clinical outcomes suggests the possibility of discontinuing antiplatelet therapy following A-TAH implantation in the future.

ASAIO JournalVol. 71(9)
University of Copenhagen (DK), Inserm (FR), Université Paris Cité (FR), Charles University (CZ), Institut Pasteur de Lille (FR), Rigshospitalet (DK), University Medical Center Utrecht (NL), Centre Hospitalier Universitaire de Lille (FR), Centre Hospitalier Universitaire de Nantes (FR), Sorbonne Paris Cité (FR), Hôpital Cochin (FR), Assistance Publique – Hôpitaux de Paris (FR), Hôpital Européen Georges-Pompidou (FR), Hôpital Pontchaillou (FR), Institute of Clinical and Experimental Medicine (CZ), Paris Cardiovascular Research Center (FR), Hôpital Saint-Louis (FR), Hôpital Laennec (FR)
Good health and well-being
Openalex Percentile: Top 34%
Mechanical Circulatory Support Devices
0.67
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.