Field-deployable cantilever-enhanced photoacoustic instrument for aerosol light absorption measurement at three wavelengths

Abstract. We demonstrate a measurement of aerosol absorption at three wavelengths for particles below 1 µm in diameter using a highly sensitive photoacoustic spectrometer. The acoustic signal is detected with a cantilever microphone, which allows for sensitive detection without the need to apply acoustic resonance to enhance the signal. The lack of resonator makes the instrument compact and well suited for field measurements. A field instrument employing the method was developed and deployed for black carbon monitoring at an air quality measurement station. The method shows excellent sensitivity for in situ aerosol absorption measurement, with detection limits of 0.016, 0.025 and 0.041 Mm−1, for simultaneous measurements at the wavelengths of 445, 520 and 638 nm, respectively, using a 1 h averaging time. The black carbon concentration measured with the new instrument is compared against filter-based photometers operating at the site, showing high correlation.

Authors

Institutions

Publication Details

Journal
Aerosol Research
Published
2025-02-28
DOI
https://doi.org/10.5194/ar-3-113-2025
Citations
3
Primary Topic
Atmospheric chemistry and aerosols
Type
article
Field-Weighted Citation Impact
3.90

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Field-deployable cantilever-enhanced photoacoustic instrument for aerosol light absorption measurement at three wavelengths

Tuomas Hieta, Aki Virkkula, Juho Karhu, Tommi Mikkonen et al.
3 citations
Aerosol Research
Atmospheric chemistry and aerosols
3.90
article

Field-deployable cantilever-enhanced photoacoustic instrument for aerosol light absorption measurement at three wavelengths

Tuomas Hieta, Aki Virkkula, Juho Karhu, Tommi Mikkonen, Hilkka Timonen, Erkki Ikonen, Markku Vainio, Joel Kuula
article en
3 citations

Abstract

Abstract. We demonstrate a measurement of aerosol absorption at three wavelengths for particles below 1 µm in diameter using a highly sensitive photoacoustic spectrometer. The acoustic signal is detected with a cantilever microphone, which allows for sensitive detection without the need to apply acoustic resonance to enhance the signal. The lack of resonator makes the instrument compact and well suited for field measurements. A field instrument employing the method was developed and deployed for black carbon monitoring at an air quality measurement station. The method shows excellent sensitivity for in situ aerosol absorption measurement, with detection limits of 0.016, 0.025 and 0.041 Mm−1, for simultaneous measurements at the wavelengths of 445, 520 and 638 nm, respectively, using a 1 h averaging time. The black carbon concentration measured with the new instrument is compared against filter-based photometers operating at the site, showing high correlation.

Aerosol ResearchVol. 3(1)
Finnish Meteorological Institute (FI), University of Helsinki (FI), Gasera (Finland) (FI), Tampere University (FI), VTT Technical Research Centre of Finland (FI), Aalto University (FI)
Academy of Finland, Jane ja Aatos Erkon Säätiö, Business Finland
Openalex Percentile: Top 8%
Atmospheric chemistry and aerosols
3.90
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.