Typical Diagnostic Reference Levels of Radiation Exposure on Neonates Under 1 kg in Mobile Chest Imaging in Incubators

The purpose of this study is to determine the typical diagnostic reference levels (DRLs) of radiation exposure values for chest radiographs in neonates (<1 kg) in mobile imaging at a University Hospital in Greece and compare these values with the existing DRL values from the literature. Patient and dosimetry data, including age, sex, weight, tube voltage (kV), tube current (mA), exposure time (s), exposure index of a digital detector (S), and dose area product (DAP) were obtained from a total of 80 chest radiography examinations performed on neonates (<1 kg and <30 days old). All examinations were performed in a single X-ray system, and all data (demographic and dosimetry data) were collected from the PACS of the hospital. Typical radiation exposure values were determined as the median value of DAP and ESD distribution. Afterward, these typical values were compared with DRL values from other countries. Three radiologists reviewed the images to evaluate image quality for dose optimization in neonatal chest radiography. From all examinations, the mean value and standard deviation of DAP was 0.13 ± 0.11 dGy·cm<sup>2</sup> (range: 0.01-0.46 dGy·cm<sup>2</sup>), and ESD was measured at 11.55 ± 4.96 μGy (range: 4.01-30.4 μGy). The typical values in terms of DAP and ESD were estimated to be 0.08 dGy·cm<sup>2</sup> and 9.87 μGy, respectively. The results show that the DAP value decreases as the exposure index increases. This study's typical values were lower than the DRLs reported in the literature because our population had lower weight and age. From the subjective evaluation of image quality, it was revealed that the vast majority of radiographs (over 80%) met the criteria for being diagnostic as they received an excellent rating in terms of noise levels, contrast, and sharpness. This study contributes to the recording of typical dose values in a sensitive and rare category of patients (neonates weighing <1 kg) as well as information on the image quality of chest X-rays that were performed in this group.

Authors

Institutions

Publication Details

Journal
Journal of Imaging
Published
2025-02-28
DOI
https://doi.org/10.3390/jimaging11030074
Primary Topic
Radiation Dose and Imaging
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Typical Diagnostic Reference Levels of Radiation Exposure on Neonates Under 1 kg in Mobile Chest Imaging in Incubators

Efstathios Efstathopoulos, Ioannis Antonakos, Maria-Eleni Zachou, George Christopoulos et al.
Journal of Imaging
Radiation Dose and Imaging
article

Typical Diagnostic Reference Levels of Radiation Exposure on Neonates Under 1 kg in Mobile Chest Imaging in Incubators

Efstathios Efstathopoulos, Ioannis Antonakos, Maria-Eleni Zachou, George Christopoulos, Matina Patsioti
article en

Abstract

The purpose of this study is to determine the typical diagnostic reference levels (DRLs) of radiation exposure values for chest radiographs in neonates (<1 kg) in mobile imaging at a University Hospital in Greece and compare these values with the existing DRL values from the literature. Patient and dosimetry data, including age, sex, weight, tube voltage (kV), tube current (mA), exposure time (s), exposure index of a digital detector (S), and dose area product (DAP) were obtained from a total of 80 chest radiography examinations performed on neonates (<1 kg and <30 days old). All examinations were performed in a single X-ray system, and all data (demographic and dosimetry data) were collected from the PACS of the hospital. Typical radiation exposure values were determined as the median value of DAP and ESD distribution. Afterward, these typical values were compared with DRL values from other countries. Three radiologists reviewed the images to evaluate image quality for dose optimization in neonatal chest radiography. From all examinations, the mean value and standard deviation of DAP was 0.13 ± 0.11 dGy·cm<sup>2</sup> (range: 0.01-0.46 dGy·cm<sup>2</sup>), and ESD was measured at 11.55 ± 4.96 μGy (range: 4.01-30.4 μGy). The typical values in terms of DAP and ESD were estimated to be 0.08 dGy·cm<sup>2</sup> and 9.87 μGy, respectively. The results show that the DAP value decreases as the exposure index increases. This study's typical values were lower than the DRLs reported in the literature because our population had lower weight and age. From the subjective evaluation of image quality, it was revealed that the vast majority of radiographs (over 80%) met the criteria for being diagnostic as they received an excellent rating in terms of noise levels, contrast, and sharpness. This study contributes to the recording of typical dose values in a sensitive and rare category of patients (neonates weighing <1 kg) as well as information on the image quality of chest X-rays that were performed in this group.

Journal of ImagingVol. 11(3)
National and Kapodistrian University of Athens (GR)
Openalex Percentile: Top 94%
Radiation Dose and Imaging
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.