Digital Ion Trap Mass Spectrometer Measurement System
This paper builds a portable digital ion trap mass spectrometer prototype by designing the vacuum system, control circuit system, and host computer of the ion trap mass spectrometer. Dextromethorphan and ketamine samples were tested. At a scanning rate of 5000Th/s, the resolution of the dextromethorphan sample reached 0.26Th, and the resolution of the ketamine sample reached 0.33Th. The actual mass spectrum collected by the prototype is basically consistent with the theoretical mass spectrum.
Authors
- Xin Wang (ORCID: https://orcid.org/0000-0003-4099-4268)
- Yong Zuo
- Xianguang Fan (ORCID: https://orcid.org/0000-0001-9787-5230)
- Li Tu (ORCID: https://orcid.org/0000-0003-3307-5554)
- Yingjie Xu (ORCID: https://orcid.org/0000-0003-2462-1840)
Institutions
- Xiamen University (CN)
- Jiangsu Institute of Metrology (CN)
- Changcheng Institute of Metrology & Measurement
Publication Details
- Published
- 2024-10-25
- DOI
- https://doi.org/10.1109/ccpqt64497.2024.00060
- Primary Topic
- Mass Spectrometry Techniques and Applications
- Type
- article
- Field-Weighted Citation Impact
- 0.00