Using smartphone pupillometer application to measure pupil size and light reflex: An unsuccessful prototype and analysis of the causes of failure

The accurate assessment of pupillary light reflex (PLR) is essential for monitoring critically ill patients, particularly those with traumatic brain injury or stroke and those in postoperative care. Smartphone-based pupillometers represent a potentially cost-effective solution for addressing this need. We developed a smartphone pupillometer application (app) and evaluated its effectiveness against the penlight test and quantitative pupillometry. This study included 50 volunteers aged >20 years and excluded individuals with neurologic or ophthalmic conditions. The app captured pupillary images by displaying a red circle on the screen, and an algorithm processed these images to calculate the pupil constriction percentage (PCP). The results revealed that the smartphone app often required multiple attempts for successful image acquisition. The obtained PCPs were consistently smaller and less variable than those obtained using the penlight test and a commercial pupillometer (app vs penlight for the right eye: 27.0% [27.0%-8.0%] vs 33.0% [32.3%-39.3%]; app vs pupillometer for the right eye: 27.0% [27.0%-28.0%] vs 35.0% [31.8%-38.3%]; app vs penlight for the left eye: 29.0% [28.0%-29.0%] vs 33.0% [29.8%-34.3%]; app vs pupillometer for the left eye: 29.0% [28.0%-29.0%] vs 36.0% [30.8%-38.0%]; P <.001 for all). Notably, the penlight and the pupillometer exhibited comparable PCPs (right eye: penlight vs pupillometer: 33.0% [32.3%-39.3%] vs 35.0% [31.8%-38.3%], P = .469; left eye: penlight vs pupillometer: 33.0% [29.8%-34.3%] vs 36.0% [30.8%-38.0%], P = .148). The app requires further refinement to yield results comparable to those of established methods. Future iterations can include alternative measurement strategies and dynamic assessment. Penlight and quantitative pupillometry remain indispensable as established tools for PLR.

Authors

Institutions

Publication Details

Journal
Medicine
Published
2025-02-28
DOI
https://doi.org/10.1097/md.0000000000041682
Citations
3
Primary Topic
Glaucoma and retinal disorders
Type
article
Field-Weighted Citation Impact
5.12
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Using smartphone pupillometer application to measure pupil size and light reflex: An unsuccessful prototype and analysis of the causes of failure

Kuo-Chang Lee, Yi‐Rong Chen, Rih-Shen Ke, Yu-Chin Huang et al.
3 citations
Medicine
Glaucoma and retinal disorders
5.12
article

Using smartphone pupillometer application to measure pupil size and light reflex: An unsuccessful prototype and analysis of the causes of failure

Kuo-Chang Lee, Yi‐Rong Chen, Rih-Shen Ke, Yu-Chin Huang, Hongwei Tan, Gwo-Jiun Horng, Kuo-Tai Chen
article en
3 citations

Abstract

The accurate assessment of pupillary light reflex (PLR) is essential for monitoring critically ill patients, particularly those with traumatic brain injury or stroke and those in postoperative care. Smartphone-based pupillometers represent a potentially cost-effective solution for addressing this need. We developed a smartphone pupillometer application (app) and evaluated its effectiveness against the penlight test and quantitative pupillometry. This study included 50 volunteers aged >20 years and excluded individuals with neurologic or ophthalmic conditions. The app captured pupillary images by displaying a red circle on the screen, and an algorithm processed these images to calculate the pupil constriction percentage (PCP). The results revealed that the smartphone app often required multiple attempts for successful image acquisition. The obtained PCPs were consistently smaller and less variable than those obtained using the penlight test and a commercial pupillometer (app vs penlight for the right eye: 27.0% [27.0%-8.0%] vs 33.0% [32.3%-39.3%]; app vs pupillometer for the right eye: 27.0% [27.0%-28.0%] vs 35.0% [31.8%-38.3%]; app vs penlight for the left eye: 29.0% [28.0%-29.0%] vs 33.0% [29.8%-34.3%]; app vs pupillometer for the left eye: 29.0% [28.0%-29.0%] vs 36.0% [30.8%-38.0%]; P <.001 for all). Notably, the penlight and the pupillometer exhibited comparable PCPs (right eye: penlight vs pupillometer: 33.0% [32.3%-39.3%] vs 35.0% [31.8%-38.3%], P = .469; left eye: penlight vs pupillometer: 33.0% [29.8%-34.3%] vs 36.0% [30.8%-38.0%], P = .148). The app requires further refinement to yield results comparable to those of established methods. Future iterations can include alternative measurement strategies and dynamic assessment. Penlight and quantitative pupillometry remain indispensable as established tools for PLR.

MedicineVol. 104(9)
Chi Mei Medical Center (TW), Southern Taiwan University of Science and Technology (TW)
Openalex Percentile: Top 6%
Glaucoma and retinal disorders
5.12
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.