The short-wave infrared (SWIR) grating imaging spectrometer based on indium gallium arsenide (InGaAs) material inverts the atmospheric methane concentration by measuring the scattered light signals in the sky. This study proposes spectral and radiometric calibration methods for the characteristics of the spectrometer, such as the small-area array, high signal-to-noise ratio, and high spectral resolution. Four spectral response function models, namely, the Gauss, Lorentz, Voigt and super-Gaussian models, were compared during spectral calibration. With a fitting residual of 0.032, the Gauss model was found to be the most suitable spectral response function for the spectrometer. Based on the spectral response function, the spectral range and spectral resolution of the spectrometer were determined to be 1592.4–1677.2 and 0.1867 nm, respectively. In addition, radiometric calibration of the spectrometer was achieved by combining an integrating sphere and linear measuring instrument. Moreover, absolute and relative radiometric calibrations of the spectrometer were performed. The low signal response problem caused by the quantum efficiency of the detector at long wavelength was corrected, and the uncertainty and non-stability uncertainty of absolute radiometric calibration were calculated to be less than 0.2%. Finally, the calibrated spectrometer was used to accurately measure the solar scattering spectrum in the SWIR band, and the solar spectrum was simulated by the radiative transfer model for verification; the measurement error was found to be 5%. Concurrently, a methane sample gas experiment was performed using the integrating-sphere light source, and the measurement error was less than 4%. This fully proves the effectiveness of the spectral and radiometric calibrations of the SWIR spectrometer and strongly guarantees a subsequent, rapid and accurate inversion of atmospheric methane concentration.

Authors

Institutions

Publication Details

Journal
Remote Sensing
Published
2025-02-28
DOI
https://doi.org/10.3390/rs17050851
Citations
1
Primary Topic
Atmospheric and Environmental Gas Dynamics
Type
article
Field-Weighted Citation Impact
1.11

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Calibration of Short-Wave Infrared Spectrometer for Atmosphere Methane Monitoring

Yi Zeng, Fenglei Liu, Liang Xi, Fuqi Si et al.
1 citations
Remote Sensing
Atmospheric and Environmental Gas Dynamics
1.11
article

Calibration of Short-Wave Infrared Spectrometer for Atmosphere Methane Monitoring

Yi Zeng, Fenglei Liu, Liang Xi, Fuqi Si, Yunkun Han, Haoran Li, Jiang Yu, Fang Lin, Kaili Wu
article en
1 citations

Abstract

The short-wave infrared (SWIR) grating imaging spectrometer based on indium gallium arsenide (InGaAs) material inverts the atmospheric methane concentration by measuring the scattered light signals in the sky. This study proposes spectral and radiometric calibration methods for the characteristics of the spectrometer, such as the small-area array, high signal-to-noise ratio, and high spectral resolution. Four spectral response function models, namely, the Gauss, Lorentz, Voigt and super-Gaussian models, were compared during spectral calibration. With a fitting residual of 0.032, the Gauss model was found to be the most suitable spectral response function for the spectrometer. Based on the spectral response function, the spectral range and spectral resolution of the spectrometer were determined to be 1592.4–1677.2 and 0.1867 nm, respectively. In addition, radiometric calibration of the spectrometer was achieved by combining an integrating sphere and linear measuring instrument. Moreover, absolute and relative radiometric calibrations of the spectrometer were performed. The low signal response problem caused by the quantum efficiency of the detector at long wavelength was corrected, and the uncertainty and non-stability uncertainty of absolute radiometric calibration were calculated to be less than 0.2%. Finally, the calibrated spectrometer was used to accurately measure the solar scattering spectrum in the SWIR band, and the solar spectrum was simulated by the radiative transfer model for verification; the measurement error was found to be 5%. Concurrently, a methane sample gas experiment was performed using the integrating-sphere light source, and the measurement error was less than 4%. This fully proves the effectiveness of the spectral and radiometric calibrations of the SWIR spectrometer and strongly guarantees a subsequent, rapid and accurate inversion of atmospheric methane concentration.

Remote SensingVol. 17(5)
University of Science and Technology of China (CN), Chinese Academy of Sciences (CN), Hefei Institutes of Physical Science (CN), Anhui Institute of Optics and Fine Mechanics (CN)
National Key Research and Development Program of China
Openalex Percentile: Top 27%
Atmospheric and Environmental Gas Dynamics
1.11
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.