Verification of the analytical performance of the urine cannabinoids immunoassay on the Abbott Alinity ci ®

The precision and reliability of the urine cannabinoids assay on the Abbott Alinity ci® analyzer were systematically evaluated. The study was conducted over 30 days in the biochemistry laboratory of Mohammed VI University Hospital. Two phases were carried out: reproducibility assessment involving daily measurements across low and high cannabinoid levels, and repeatability testing involving 30 replicates per sample. A homogeneous immunoenzymatic approach was employed for cannabinoid quantification. Data analysis utilized the BYG middleware and adhered to the manufacturer standards. For reproducibility, the coefficient of variation (CV) values were low, ranging from 2,6% to 3,1%. Repeatability CV values were also low, varying from 2,8% to 3,7%. Results were consistent with quality control limits, confirming the assay's precision and reliability. This study highlights the robustness of the immunoenzymatic method on the Alinity ci® platform for urine cannabinoid analysis, with significant implications for accurate patient care. By adhering to stringent analytical performance verification protocols, laboratories ensure dependable and clinically meaningful outcomes. This research contributes to the foundational knowledge supporting the reliability of urine cannabinoid measurement methodologies.

Authors

Publication Details

Journal
GSC Biological and Pharmaceutical Sciences
Published
2025-02-26
DOI
https://doi.org/10.30574/gscbps.2025.30.2.0056
Primary Topic
Metabolomics and Mass Spectrometry Studies
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Verification of the analytical performance of the urine cannabinoids immunoassay on the Abbott Alinity ci ®

Dounia El Moujtahide, Mohammed Choukri, El Houcine Sebbar, Sohaib El kandouni
GSC Biological and Pharmaceutical Sciences
Metabolomics and Mass Spectrometry Studies
article

Verification of the analytical performance of the urine cannabinoids immunoassay on the Abbott Alinity ci ®

Dounia El Moujtahide, Mohammed Choukri, El Houcine Sebbar, Sohaib El kandouni
article en

Abstract

The precision and reliability of the urine cannabinoids assay on the Abbott Alinity ci® analyzer were systematically evaluated. The study was conducted over 30 days in the biochemistry laboratory of Mohammed VI University Hospital. Two phases were carried out: reproducibility assessment involving daily measurements across low and high cannabinoid levels, and repeatability testing involving 30 replicates per sample. A homogeneous immunoenzymatic approach was employed for cannabinoid quantification. Data analysis utilized the BYG middleware and adhered to the manufacturer standards. For reproducibility, the coefficient of variation (CV) values were low, ranging from 2,6% to 3,1%. Repeatability CV values were also low, varying from 2,8% to 3,7%. Results were consistent with quality control limits, confirming the assay's precision and reliability. This study highlights the robustness of the immunoenzymatic method on the Alinity ci® platform for urine cannabinoid analysis, with significant implications for accurate patient care. By adhering to stringent analytical performance verification protocols, laboratories ensure dependable and clinically meaningful outcomes. This research contributes to the foundational knowledge supporting the reliability of urine cannabinoid measurement methodologies.

GSC Biological and Pharmaceutical SciencesVol. 30(2)
Openalex Percentile: Top 97%
Metabolomics and Mass Spectrometry Studies
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.