Verification of the analytical performance of the urine ecstasy immunoassay on the Abbott Alinity ci ®

The precision and reliability of the urine cannabinoids immunoassay on the Abbott Alinity ci® analyzer were systematically evaluated over 30 days in the biochemistry laboratory of Mohammed VI University Hospital. The study consisted of two phases: reproducibility assessment, involving daily measurements at low and high cannabinoid concentrations, and repeatability testing, with 30 replicates per sample. Cannabinoid quantification was performed using a homogeneous immunoenzymatic assay, with data analysis conducted via the BYG middleware, following the manufacturer's guidelines. The coefficient of variation (CV) for reproducibility ranged from 2.6% to 3.1%, while repeatability CV values varied between 2.8% and 3.7%, all within quality control limits, confirming the assay’s high precision and reliability. These findings underscore the robustness of the Alinity ci® platform for urine cannabinoid analysis, ensuring accurate and reproducible results critical for clinical and forensic applications. By adhering to rigorous analytical performance verification protocols, laboratories can ensure dependable and clinically meaningful outcomes, reinforcing the reliability of urine cannabinoid testing methodologies.

Authors

Publication Details

Journal
GSC Biological and Pharmaceutical Sciences
Published
2025-02-26
DOI
https://doi.org/10.30574/gscbps.2025.30.2.0051
Primary Topic
Urinary Tract Infections Management
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Verification of the analytical performance of the urine ecstasy immunoassay on the Abbott Alinity ci ®

Dounia El Moujtahide, Mohammed Choukri, El Houcine Sebbar, Sohaib El kandouni
GSC Biological and Pharmaceutical Sciences
Urinary Tract Infections Management
article

Verification of the analytical performance of the urine ecstasy immunoassay on the Abbott Alinity ci ®

Dounia El Moujtahide, Mohammed Choukri, El Houcine Sebbar, Sohaib El kandouni
article en

Abstract

The precision and reliability of the urine cannabinoids immunoassay on the Abbott Alinity ci® analyzer were systematically evaluated over 30 days in the biochemistry laboratory of Mohammed VI University Hospital. The study consisted of two phases: reproducibility assessment, involving daily measurements at low and high cannabinoid concentrations, and repeatability testing, with 30 replicates per sample. Cannabinoid quantification was performed using a homogeneous immunoenzymatic assay, with data analysis conducted via the BYG middleware, following the manufacturer's guidelines. The coefficient of variation (CV) for reproducibility ranged from 2.6% to 3.1%, while repeatability CV values varied between 2.8% and 3.7%, all within quality control limits, confirming the assay’s high precision and reliability. These findings underscore the robustness of the Alinity ci® platform for urine cannabinoid analysis, ensuring accurate and reproducible results critical for clinical and forensic applications. By adhering to rigorous analytical performance verification protocols, laboratories can ensure dependable and clinically meaningful outcomes, reinforcing the reliability of urine cannabinoid testing methodologies.

GSC Biological and Pharmaceutical SciencesVol. 30(2)
Openalex Percentile: Top 94%
Urinary Tract Infections Management
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.