Diagnostic limitations in advanced stage peripheral arterial disease in a cadaveric study comparing photon-counting and energy-integrating CT detectors

To evaluate the limits of 1st-generation dual-source photon-counting detector CT (PCD-CT) and 3rd-generation dual-source energy-integrating-detector (EID-CT) regarding imaging of advanced stage peripheral arterial disease (ASPAD) of the femoral runoff. One human cadaver with ASPAD of the superficial femoral arteries was surgically prepared to establish continuous extracorporeal perfusion of the right upper leg. In addition to one stent already in place, three more stents were deployed in positions with severe calcification and stenosis to create thirteen different scenarios of ASPAD. CT angiographies with different radiation dose (CTDI<sub>vol</sub> 10, 5, 3 mGy) and matching convolution kernels were performed with PCD-CT and EID-CT. In-stent lumen visibility, signal-to-noise ratio (SNR), and luminal attenuation were assessed quantitatively. Results were compared using analyses of variance with a PCD-CT maximum dose and resolution scan (96 mGy, BV89) serving as standard of reference. Highest and lowest stent lumen visibility was observed with PCD-CT BV76 (97 ± 2%) and EID BV40 (77 ± 5%), respectively. Severe stent underexpansion in conjunction with heavy calcification resulted in the worst lumen visibility. PCD-CT displayed superior dose efficiency, yielding comparable SNR at 3 mGy to EID-CT at 10 mGy (p = 0.27). Luminal attenuation was higher for PCD-CT regardless of dose and reconstruction settings (max. 369 ± 19 HU, BV76, 5 mGy vs. 329 ± 12 HU for EID, BV59, 5 mGy; p < 0.001). PCD-CT realises substantially higher image quality than EID-CT, thereby enhancing assessment of the femoral vasculature in ASPAD. Furthermore, this indicates substantial radiation dose and contrast agent volume saving potential. Both scanners show limitations in very low luminal diameters.

Authors

Institutions

Publication Details

Journal
Scientific Reports
Published
2025-02-26
DOI
https://doi.org/10.1038/s41598-025-91239-x
Citations
2
Primary Topic
Advanced X-ray and CT Imaging
Type
article
Field-Weighted Citation Impact
1.35
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Diagnostic limitations in advanced stage peripheral arterial disease in a cadaveric study comparing photon-counting and energy-integrating CT detectors

Florian Kleefeldt, Henner Huflage, Dominik Peter, Philipp Gruschwitz et al.
2 citations
Scientific Reports
Advanced X-ray and CT Imaging
1.35
article

Diagnostic limitations in advanced stage peripheral arterial disease in a cadaveric study comparing photon-counting and energy-integrating CT detectors

Florian Kleefeldt, Henner Huflage, Dominik Peter, Philipp Gruschwitz, Süleyman Ergün, Jan‐Peter Grunz, Thorsten Alexander Bley, Anne Marie Augustin, Viktor Hartung, Sven Lichthardt
article en
2 citations

Abstract

To evaluate the limits of 1st-generation dual-source photon-counting detector CT (PCD-CT) and 3rd-generation dual-source energy-integrating-detector (EID-CT) regarding imaging of advanced stage peripheral arterial disease (ASPAD) of the femoral runoff. One human cadaver with ASPAD of the superficial femoral arteries was surgically prepared to establish continuous extracorporeal perfusion of the right upper leg. In addition to one stent already in place, three more stents were deployed in positions with severe calcification and stenosis to create thirteen different scenarios of ASPAD. CT angiographies with different radiation dose (CTDI<sub>vol</sub> 10, 5, 3 mGy) and matching convolution kernels were performed with PCD-CT and EID-CT. In-stent lumen visibility, signal-to-noise ratio (SNR), and luminal attenuation were assessed quantitatively. Results were compared using analyses of variance with a PCD-CT maximum dose and resolution scan (96 mGy, BV89) serving as standard of reference. Highest and lowest stent lumen visibility was observed with PCD-CT BV76 (97 ± 2%) and EID BV40 (77 ± 5%), respectively. Severe stent underexpansion in conjunction with heavy calcification resulted in the worst lumen visibility. PCD-CT displayed superior dose efficiency, yielding comparable SNR at 3 mGy to EID-CT at 10 mGy (p = 0.27). Luminal attenuation was higher for PCD-CT regardless of dose and reconstruction settings (max. 369 ± 19 HU, BV76, 5 mGy vs. 329 ± 12 HU for EID, BV59, 5 mGy; p < 0.001). PCD-CT realises substantially higher image quality than EID-CT, thereby enhancing assessment of the femoral vasculature in ASPAD. Furthermore, this indicates substantial radiation dose and contrast agent volume saving potential. Both scanners show limitations in very low luminal diameters.

Scientific ReportsVol. 15(1)
University of Wisconsin–Madison (US), University of Würzburg (DE), Bayreuth Medical Center (DE), Universitätsklinikum Würzburg (DE)
Affordable and clean energy
Openalex Percentile: Top 22%
Advanced X-ray and CT Imaging
1.35
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.