DiCARTTM device to measure capillary refill time: a validation study in patients with acute circulatory failure

Capillary Refill Time (CRT) is a valuable metric to assess cutaneous perfusion. Its prognostic value in patients with acute circulatory failure has been reported as improved when the measurement is standardized. The DiCART™ device is a fully automated CRT measurement tool requiring validation. We conducted a comparative interventional single-center study including 25 patients with acute circulatory failure, to evaluate the agreement between CRT measured by an automated measurement device (CRT<sub>DiCART</sub>) and CRT measured clinically (CRT<sub>CLIN</sub>). CRT was measured on the fingertip, chest, and knee. Three measurements were performed at each location to obtain an average for each site. The measurements were conducted both clinically and using the DiCART™ device by two different operators, each blinded to the results. Agreement was determined using intraclass correlation coefficient (ICC) and Bland and Altman analysis. The ICC between CRT<sub>CLIN</sub> and CRT<sub>DiCART</sub> was 0.46 (95% Confidence Interval (CI) 0.32, 0.59) across all measurement sites; the mean bias was 0.23s (95% CI -0.17, 0.64), with upper Limit of Agreement (LoA) 2.77s (95% CI 2.44, 3.20) and lower LoA - 2.30s (-2.73, -1.97). Intra observer ICC was 0.85 (95% CI 0.74, 0.91) for CRT<sub>CLIN</sub> and 0.43 (95% CI 0.15, 0.64) for CRT<sub>DICART</sub>. Inter observer ICC was 0.86 (95% CI 0.76, 0.92) for CRT<sub>CLIN</sub> and was 0.41 (95% CI 0.14, 0,63) for CRT<sub>DICART</sub>. The DiCART™ device showed poor agreement with clinical CRT in patients with acute circulatory failure, which does not support its use in routine practice.

Authors

Institutions

Publication Details

Journal
Journal of Clinical Monitoring and Computing
Published
2025-02-26
DOI
https://doi.org/10.1007/s10877-025-01271-5
Citations
2
Primary Topic
Sepsis Diagnosis and Treatment
Type
article
Field-Weighted Citation Impact
2.71

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

DiCARTTM device to measure capillary refill time: a validation study in patients with acute circulatory failure

Martin Ruste, A. Descamps, Matthias Jacquet‐Lagrèze, Jean-Luc Fellahi et al.
2 citations
Journal of Clinical Monitoring and Computing
Sepsis Diagnosis and Treatment
2.71
article

DiCARTTM device to measure capillary refill time: a validation study in patients with acute circulatory failure

Martin Ruste, A. Descamps, Matthias Jacquet‐Lagrèze, Jean-Luc Fellahi, Thomas Aussal
article en
2 citations

Abstract

Capillary Refill Time (CRT) is a valuable metric to assess cutaneous perfusion. Its prognostic value in patients with acute circulatory failure has been reported as improved when the measurement is standardized. The DiCART™ device is a fully automated CRT measurement tool requiring validation. We conducted a comparative interventional single-center study including 25 patients with acute circulatory failure, to evaluate the agreement between CRT measured by an automated measurement device (CRT<sub>DiCART</sub>) and CRT measured clinically (CRT<sub>CLIN</sub>). CRT was measured on the fingertip, chest, and knee. Three measurements were performed at each location to obtain an average for each site. The measurements were conducted both clinically and using the DiCART™ device by two different operators, each blinded to the results. Agreement was determined using intraclass correlation coefficient (ICC) and Bland and Altman analysis. The ICC between CRT<sub>CLIN</sub> and CRT<sub>DiCART</sub> was 0.46 (95% Confidence Interval (CI) 0.32, 0.59) across all measurement sites; the mean bias was 0.23s (95% CI -0.17, 0.64), with upper Limit of Agreement (LoA) 2.77s (95% CI 2.44, 3.20) and lower LoA - 2.30s (-2.73, -1.97). Intra observer ICC was 0.85 (95% CI 0.74, 0.91) for CRT<sub>CLIN</sub> and 0.43 (95% CI 0.15, 0.64) for CRT<sub>DICART</sub>. Inter observer ICC was 0.86 (95% CI 0.76, 0.92) for CRT<sub>CLIN</sub> and was 0.41 (95% CI 0.14, 0,63) for CRT<sub>DICART</sub>. The DiCART™ device showed poor agreement with clinical CRT in patients with acute circulatory failure, which does not support its use in routine practice.

Journal of Clinical Monitoring and ComputingVol. 39(5)
Université Claude Bernard Lyon 1 (FR), Inserm (FR), Lyon College (US), Hospices Civils de Lyon (FR), Laboratoire de recherche en cardiovasculaire, métabolisme, diabétologie et nutrition (FR), Hôpital Louis Pradel (FR)
Hospices Civils de Lyon
No poverty
Openalex Percentile: Top 12%
Sepsis Diagnosis and Treatment
2.71
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.