Viscosity measurement of small volume liquid samples in a falling body viscometer employing compact electromagnetic actuation and sensing

Falling body (FB) methods have unique advantages in the viscosity measurement of high-temperature and high-pressure liquids because of the durability of the sensing parts and their suitability for carrying out measurements in a metal closed cylindrical chamber (CCC). However, the current methods usually use two coils fixed away from each other to sense the position and speed of the FB, which leads to a long travel distance of the FB, a large CCC, and the inability of the FB to automatically reset. This paper proposes a novel compact liquid viscosity measurement device using only one coil outside a stainless steel CCC to sense the moving speed of the iron FB and uses another coil fixed close to the sensing coil to reset the FB. Both the FB travel distance and the CCC size are very small. The stable velocity of the FB is directly measured according to the linear relationship between the impedance of the sensing coil and the position of the FB, and then the velocity is used to calculate the liquid viscosity. Viscosity tests were conducted on four standard viscosity liquids at different temperatures. The results show that for liquids with a viscosity range of 4.417 mPa·s∼88.44 mPa·s, the descent time of FBs is between 2 and 19 s. The relative error is concentrated within 5 %, and the standard deviation falls between 0.01 mPa·s and 1.46 mPa·s. Within the 95 % confidence interval ( k = 2), the expanded uncertainty of the four liquids viscosity measurement at different temperatures ranges from 0.74 mPa·s to 2.93 mPa·s, and the relative expanded uncertainty ranges from 6.32 % to 10.6 %. The measurement results have high accuracy and repeatability. Fig. 1 Viscosity measurement apparatus • A novel compact device for measuring liquid viscosity in a closed cylindrical chamber is proposed. • Velocity of falling body(FB) is proportional to coil impedance’s change rate and inversely proportional to liquid viscosity. • The FB's travel distance is short, so the chamber is small and the FB can be reset by another coil fixed close. • The relative error is concentrated within 5 %, and the standard deviation is between 0.01mPa·s-1.46mPa·s.

Authors

Publication Details

Journal
Sensors and Actuators A Physical
Published
2025-02-26
DOI
https://doi.org/10.1016/j.sna.2025.116401
Citations
1
Primary Topic
Microfluidic and Capillary Electrophoresis Applications
Type
article
Field-Weighted Citation Impact
0.67

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Viscosity measurement of small volume liquid samples in a falling body viscometer employing compact electromagnetic actuation and sensing

Xinjing Huang, Kun Zhang, Jian Li, Hongbin Zhang et al.
1 citations
Sensors and Actuators A Physical
Microfluidic and Capillary Electrophoresis Applications
0.67
article

Viscosity measurement of small volume liquid samples in a falling body viscometer employing compact electromagnetic actuation and sensing

Xinjing Huang, Kun Zhang, Jian Li, Hongbin Zhang, Jinyu Ma, Xu Bian
article en
1 citations

Abstract

Falling body (FB) methods have unique advantages in the viscosity measurement of high-temperature and high-pressure liquids because of the durability of the sensing parts and their suitability for carrying out measurements in a metal closed cylindrical chamber (CCC). However, the current methods usually use two coils fixed away from each other to sense the position and speed of the FB, which leads to a long travel distance of the FB, a large CCC, and the inability of the FB to automatically reset. This paper proposes a novel compact liquid viscosity measurement device using only one coil outside a stainless steel CCC to sense the moving speed of the iron FB and uses another coil fixed close to the sensing coil to reset the FB. Both the FB travel distance and the CCC size are very small. The stable velocity of the FB is directly measured according to the linear relationship between the impedance of the sensing coil and the position of the FB, and then the velocity is used to calculate the liquid viscosity. Viscosity tests were conducted on four standard viscosity liquids at different temperatures. The results show that for liquids with a viscosity range of 4.417 mPa·s∼88.44 mPa·s, the descent time of FBs is between 2 and 19 s. The relative error is concentrated within 5 %, and the standard deviation falls between 0.01 mPa·s and 1.46 mPa·s. Within the 95 % confidence interval ( k = 2), the expanded uncertainty of the four liquids viscosity measurement at different temperatures ranges from 0.74 mPa·s to 2.93 mPa·s, and the relative expanded uncertainty ranges from 6.32 % to 10.6 %. The measurement results have high accuracy and repeatability. Fig. 1 Viscosity measurement apparatus • A novel compact device for measuring liquid viscosity in a closed cylindrical chamber is proposed. • Velocity of falling body(FB) is proportional to coil impedance’s change rate and inversely proportional to liquid viscosity. • The FB's travel distance is short, so the chamber is small and the FB can be reset by another coil fixed close. • The relative error is concentrated within 5 %, and the standard deviation is between 0.01mPa·s-1.46mPa·s.

Sensors and Actuators A PhysicalVol. 387
National Natural Science Foundation of China, Guangxi Key Laboratory of Automatic Detection Technology and Instrument Foundation, Guangxi Key Laboratory of Automatic Detecting Technology and Instruments, Guilin University of Electronic Technology
Affordable and clean energy
Openalex Percentile: Top 34%
Microfluidic and Capillary Electrophoresis Applications
0.67
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.