3D<i>π</i>: three-dimensional positron imaging, a novel total-body PET scanner using xenon-doped liquid argon scintillator

<i>Objective.</i>This paper introduces a novel PET imaging methodology called 3-dimensional positron imaging (3D<i>π</i>), which integrates total-body coverage, time-of-flight (TOF) technology, ultra-low dose imaging capabilities, and ultra-fast readout electronics inspired by emerging technology from the DarkSide collaboration.<i>Approach.</i>The study evaluates the performance of 3D<i>π</i>using Monte Carlo simulations based on NEMA NU 2-2018 protocols. The methodology employs a homogenous, monolithic scintillator composed of liquid argon (LAr) doped with xenon (Xe) with silicon photomultipliers (SiPMs) operating at cryogenic temperatures.<i>Main results.</i>Substantial improvements in system performance are observed, with the 3D<i>π</i>system achieving a noise equivalent count rate of 3.2 Mcps at 17.3 kBq ml<sup>-1</sup>, continuing to increase up to 4.3 Mcps at 40 kBq ml<sup>-1</sup>. Spatial resolution measurements show an average FWHM of 2.7 mm across both axial positions. The system exhibits superior sensitivity, with values reaching 373 kcps MBq<sup>-1</sup>with a line source at the center of the field of view. Additionally, 3D<i>π</i>achieves a TOF resolution of 151 ps at 5.3 kBq ml<sup>-1</sup>, highlighting its potential to produce high-quality images with reduced noise levels.<i>Significance.</i>The study underscores the potential of 3D<i>π</i>in improving PET imaging performance, offering the potential for shorter scan times and reduced radiation exposure for patients. The Xe-doped LAr offers advantages such as fast scintillation, enhanced light yield, and cost-effectiveness. Future research will focus on optimizing system geometry and further refining reconstruction algorithms to exploit the strengths of 3D<i>π</i>for clinical applications.

Authors

Institutions

Publication Details

Journal
Physics in Medicine and Biology
Published
2025-02-26
DOI
https://doi.org/10.1088/1361-6560/adbaac
Citations
1
Primary Topic
Medical Imaging Techniques and Applications
Type
article
Field-Weighted Citation Impact
1.36

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

3Dπ: three-dimensional positron imaging, a novel total-body PET scanner using xenon-doped liquid argon scintillator

Azam Zabihi, M. Laí, M. Rolo, I. Ahmad et al.
1 citations
Physics in Medicine and Biology
Medical Imaging Techniques and Applications
1.36
article

3Dπ: three-dimensional positron imaging, a novel total-body PET scanner using xenon-doped liquid argon scintillator

Azam Zabihi, M. Laí, M. Rolo, I. Ahmad, D. Franco, C. Galbiati, M. Wada, Xinran Li, D. Marlow, F. Gabriele, A. Renshaw, S. Westerdale, A. P. Ramirez
article en
1 citations

Abstract

<i>Objective.</i>This paper introduces a novel PET imaging methodology called 3-dimensional positron imaging (3D<i>π</i>), which integrates total-body coverage, time-of-flight (TOF) technology, ultra-low dose imaging capabilities, and ultra-fast readout electronics inspired by emerging technology from the DarkSide collaboration.<i>Approach.</i>The study evaluates the performance of 3D<i>π</i>using Monte Carlo simulations based on NEMA NU 2-2018 protocols. The methodology employs a homogenous, monolithic scintillator composed of liquid argon (LAr) doped with xenon (Xe) with silicon photomultipliers (SiPMs) operating at cryogenic temperatures.<i>Main results.</i>Substantial improvements in system performance are observed, with the 3D<i>π</i>system achieving a noise equivalent count rate of 3.2 Mcps at 17.3 kBq ml<sup>-1</sup>, continuing to increase up to 4.3 Mcps at 40 kBq ml<sup>-1</sup>. Spatial resolution measurements show an average FWHM of 2.7 mm across both axial positions. The system exhibits superior sensitivity, with values reaching 373 kcps MBq<sup>-1</sup>with a line source at the center of the field of view. Additionally, 3D<i>π</i>achieves a TOF resolution of 151 ps at 5.3 kBq ml<sup>-1</sup>, highlighting its potential to produce high-quality images with reduced noise levels.<i>Significance.</i>The study underscores the potential of 3D<i>π</i>in improving PET imaging performance, offering the potential for shorter scan times and reduced radiation exposure for patients. The Xe-doped LAr offers advantages such as fast scintillation, enhanced light yield, and cost-effectiveness. Future research will focus on optimizing system geometry and further refining reconstruction algorithms to exploit the strengths of 3D<i>π</i>for clinical applications.

Physics in Medicine and BiologyVol. 70(6)
University of California, Riverside (US), Centre National de la Recherche Scientifique (FR), Lawrence Berkeley National Laboratory (US), University of California, Los Angeles (US), University of Cagliari (IT), Princeton University (US), Université Paris Cité (FR), Laboratoire AstroParticule et Cosmologie (FR), Istituto Nazionale di Fisica Nucleare, Sezione di Torino (IT), Nicolaus Copernicus Astronomical Center (PL), Gran Sasso Science Institute (IT), University of Houston (US), Polish Academy of Sciences (PL)
Division of Computing and Communication Foundations
Industry, innovation and infrastructure
Openalex Percentile: Top 23%
Medical Imaging Techniques and Applications
1.36
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.