A Development of an Embedded System Software for IMU Sensor Data Logging During Clinical Assessment by Using 6-Minute Walk Test

In the traditional 6-minute walk test (6MWT), doctors or physical therapists manually record data while simultaneously monitoring the patient. This research aims to develop software to control an embedded system for recording movement data during the 6MWT using IMU sensors. The system will incorporate FreeRTOS (Free Real-Time Operating System) to efficiently manage multitasking operations. The device is designed to record walking data for 6 minutes and store it on an SD card in. csv format at a sampling rate of 100 Hz. Experimental results show that the developed device can consistently record walking data at 100 Hz without data loss during the 6MWT.

Authors

Institutions

Publication Details

Published
2024-11-21
DOI
https://doi.org/10.1109/bmeicon64021.2024.10896272
Primary Topic
Wireless Body Area Networks
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

A Development of an Embedded System Software for IMU Sensor Data Logging During Clinical Assessment by Using 6-Minute Walk Test

Weerayot Aramphianlert, Supanun Chitmeta, Phusanisa Premying, Chamaiporn Sikjamsri
Wireless Body Area Networks
article

A Development of an Embedded System Software for IMU Sensor Data Logging During Clinical Assessment by Using 6-Minute Walk Test

Weerayot Aramphianlert, Supanun Chitmeta, Phusanisa Premying, Chamaiporn Sikjamsri
article en

Abstract

In the traditional 6-minute walk test (6MWT), doctors or physical therapists manually record data while simultaneously monitoring the patient. This research aims to develop software to control an embedded system for recording movement data during the 6MWT using IMU sensors. The system will incorporate FreeRTOS (Free Real-Time Operating System) to efficiently manage multitasking operations. The device is designed to record walking data for 6 minutes and store it on an SD card in. csv format at a sampling rate of 100 Hz. Experimental results show that the developed device can consistently record walking data at 100 Hz without data loss during the 6MWT.

Srinakharinwirot University (TH)
Openalex Percentile: Top 75%
Wireless Body Area Networks
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

A Development of an Embedded System Software for IMU Sensor Data Logging During Clinical Assessment by Using 6-Minute Walk Test — Weerayot Aramphianlert, Supanun Chitmeta, et al. · (2024) | TGRS Research Map | TGRS