Evaluating use of laser ablation quadrupole inductively coupled plasma mass spectrometry for low concentration lead isotope measurements

Lead (Pb) concentrations and isotopic compositions in environmental samples can provide important insights into Pb sources, pathways, and distributions. However, low Pb levels in many environmental samples present challenges for precise Pb isotopic measurements. This study evaluates the use of laser ablation quadrupole inductively coupled plasma mass spectrometry (LA-Q-ICP-MS) combined with collisional focusing and an Aerosol Rapid Introduction System (ARIS) for measuring Pb isotopes in reference materials with variable Pb concentrations (0.3–11 μg g −1 ) and environmental samples (i.e., Pacific oyster shells) with Pb concentrations below 1 μg g −1 . Accuracy and precision were assessed using three glass reference materials (BCR-2G, BHVO-2G, and GOR128-G). For 206 Pb/ 207 Pb and 208 Pb/ 207 Pb ratios, measurements were within 0.2 % and 2.8 % of reported values for all reference materials, with decreased accuracy associated with low count rates. The precision was 0.4%–1.8 % relative standard deviation (RSD) for certified reference materials. Analytical uncertainty was lower for reference materials (∼1.2 % RSD for 206 Pb/ 207 Pb and ∼1.3 % RSD for 208 Pb/ 207 Pb) compared to oyster shells (∼3.6 % RSD for 206 Pb/ 207 Pb and ∼3.8 % RSD for 208 Pb/ 207 Pb), likely due to low Pb concentrations and sample heterogeneity. While Pb isotopic measurements demonstrated good repeatability across parallel transects of the same oyster shell (r = 0.7), the higher analytical uncertainties (>0.5 % RSD) limit reliable source identification or apportionment in these natural samples. In approaching new studies with the methods presented here, researchers will need to determine if for a given application the uncertainties on measured sample Pb isotope ratios are sufficiently small. • ARIS, LA-Q-ICP-MS and collisional focusing are promising for isotopic analyses. • Higher concentrations yielded improved accuracy and uncertainty in standards. • Heterogenous cross sections are repeatable in natural samples. • Uncertainty limits applications in low concentration natural samples.

Authors

Institutions

Publication Details

Journal
International Journal of Mass Spectrometry
Published
2025-02-26
DOI
https://doi.org/10.1016/j.ijms.2025.117429
Primary Topic
Isotope Analysis in Ecology
Type
article
Field-Weighted Citation Impact
0.00

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Evaluating use of laser ablation quadrupole inductively coupled plasma mass spectrometry for low concentration lead isotope measurements

Adam J.R. Kent, Alyssa E. Shiel, Sophia M. Wensman, Chris Russo
International Journal of Mass Spectrometry
Isotope Analysis in Ecology
article

Evaluating use of laser ablation quadrupole inductively coupled plasma mass spectrometry for low concentration lead isotope measurements

Adam J.R. Kent, Alyssa E. Shiel, Sophia M. Wensman, Chris Russo
article en

Abstract

Lead (Pb) concentrations and isotopic compositions in environmental samples can provide important insights into Pb sources, pathways, and distributions. However, low Pb levels in many environmental samples present challenges for precise Pb isotopic measurements. This study evaluates the use of laser ablation quadrupole inductively coupled plasma mass spectrometry (LA-Q-ICP-MS) combined with collisional focusing and an Aerosol Rapid Introduction System (ARIS) for measuring Pb isotopes in reference materials with variable Pb concentrations (0.3–11 μg g −1 ) and environmental samples (i.e., Pacific oyster shells) with Pb concentrations below 1 μg g −1 . Accuracy and precision were assessed using three glass reference materials (BCR-2G, BHVO-2G, and GOR128-G). For 206 Pb/ 207 Pb and 208 Pb/ 207 Pb ratios, measurements were within 0.2 % and 2.8 % of reported values for all reference materials, with decreased accuracy associated with low count rates. The precision was 0.4%–1.8 % relative standard deviation (RSD) for certified reference materials. Analytical uncertainty was lower for reference materials (∼1.2 % RSD for 206 Pb/ 207 Pb and ∼1.3 % RSD for 208 Pb/ 207 Pb) compared to oyster shells (∼3.6 % RSD for 206 Pb/ 207 Pb and ∼3.8 % RSD for 208 Pb/ 207 Pb), likely due to low Pb concentrations and sample heterogeneity. While Pb isotopic measurements demonstrated good repeatability across parallel transects of the same oyster shell (r = 0.7), the higher analytical uncertainties (>0.5 % RSD) limit reliable source identification or apportionment in these natural samples. In approaching new studies with the methods presented here, researchers will need to determine if for a given application the uncertainties on measured sample Pb isotope ratios are sufficiently small. • ARIS, LA-Q-ICP-MS and collisional focusing are promising for isotopic analyses. • Higher concentrations yielded improved accuracy and uncertainty in standards. • Heterogenous cross sections are repeatable in natural samples. • Uncertainty limits applications in low concentration natural samples.

International Journal of Mass SpectrometryVol. 511
Oregon State University (US), Desert Research Institute (US)
Oregon Sea Grant, Oregon State University
Openalex Percentile: Top 97%
Isotope Analysis in Ecology
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.