This study evaluated hybrid iterative reconstruction in ultra-low-dose computed tomography (ULDCT) for solid pulmonary nodule detection. A 256-slice CT machine operating at 120 kVp imaged a chest phantom with 5 mm nodules. The imaging process involved adjusting low-dose computed tomography (LDCT) settings and conducting 3 ULDCT scans (A-C) with varied minimum and maximum mA settings (10/40 mA). Images were processed using iDose4 iterative reconstruction at levels 5 to 7. Measurements were taken for noise, signal-to-noise ratio (SNR), contrast-to-noise ratio (CNR), noise power spectrum (NPS), and detectability index (D') to assess image quality, noise texture, and detectability. Analysis of variance (ANOVA) was used to compare the protocols. Noise levels varied significantly across iDose4 iterative reconstruction levels, with the highest noise at 178 HU in iDose4 L5 (protocol C) and the lowest at 54.85 HU in level 7 (protocol A). ULDCT scans showed noise increases of 38.5%, 104.2%, and 118.7% for protocols A, B, and C, respectively, compared to LDCT. Protocol A (iDose4 level 7) significantly improved SNR and CNR (P < .001). The mean volume CT dose index was 2.4 mGy for LDCT and 2.0 mGy, 1.2 mGy, and 0.7 mGy for ULDCT protocols A, B, and C, respectively. Increasing iDose4 levels reduced noise magnitude in the NPS and improved the D'. ULDCT with iDose4 level 7 provides diagnostically acceptable image quality for solid pulmonary nodule assessment at significantly reduced radiation doses. This approach, supported by advanced metrics like NPS and D', demonstrates a potential pathway for safer, effective lung cancer screening in high-risk populations. Further clinical studies are needed to validate these findings in diverse patient populations.

Authors

Institutions

Publication Details

Journal
Medicine
Published
2025-02-21
DOI
https://doi.org/10.1097/md.0000000000041612
Primary Topic
Radiation Dose and Imaging
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Hybrid iterative reconstruction in ultra-low-dose CT for accurate pulmonary nodule assessment: A Phantom study

Li‐Chuan Huang, Pingan Wu, Liguo Chen, Hung‐Wen Kao et al.
Medicine
Radiation Dose and Imaging
article

Hybrid iterative reconstruction in ultra-low-dose CT for accurate pulmonary nodule assessment: A Phantom study

Li‐Chuan Huang, Pingan Wu, Liguo Chen, Hung‐Wen Kao, Hsing‐Yang Tu, Ming‐Huei Sheu
article en

Abstract

This study evaluated hybrid iterative reconstruction in ultra-low-dose computed tomography (ULDCT) for solid pulmonary nodule detection. A 256-slice CT machine operating at 120 kVp imaged a chest phantom with 5 mm nodules. The imaging process involved adjusting low-dose computed tomography (LDCT) settings and conducting 3 ULDCT scans (A-C) with varied minimum and maximum mA settings (10/40 mA). Images were processed using iDose4 iterative reconstruction at levels 5 to 7. Measurements were taken for noise, signal-to-noise ratio (SNR), contrast-to-noise ratio (CNR), noise power spectrum (NPS), and detectability index (D') to assess image quality, noise texture, and detectability. Analysis of variance (ANOVA) was used to compare the protocols. Noise levels varied significantly across iDose4 iterative reconstruction levels, with the highest noise at 178 HU in iDose4 L5 (protocol C) and the lowest at 54.85 HU in level 7 (protocol A). ULDCT scans showed noise increases of 38.5%, 104.2%, and 118.7% for protocols A, B, and C, respectively, compared to LDCT. Protocol A (iDose4 level 7) significantly improved SNR and CNR (P < .001). The mean volume CT dose index was 2.4 mGy for LDCT and 2.0 mGy, 1.2 mGy, and 0.7 mGy for ULDCT protocols A, B, and C, respectively. Increasing iDose4 levels reduced noise magnitude in the NPS and improved the D'. ULDCT with iDose4 level 7 provides diagnostically acceptable image quality for solid pulmonary nodule assessment at significantly reduced radiation doses. This approach, supported by advanced metrics like NPS and D', demonstrates a potential pathway for safer, effective lung cancer screening in high-risk populations. Further clinical studies are needed to validate these findings in diverse patient populations.

MedicineVol. 104(8)
Tzu Chi University (TW)
Good health and well-being
Openalex Percentile: Top 95%
Radiation Dose and Imaging
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.