Enhancing the precision of impedance measurement from 5 kHz to 1 MHz through self-identification of parasitic parameters

<i>Objective.</i>Electrical impedance tomography (EIT) generates cross-sectional images through non-invasive impedance measurements from surface electrodes. While impedance above 200 kHz can reveal intracellular properties, most existing EIT images are published at frequencies below 200 kHz. When frequencies exceed 200 kHz, the accuracy of impedance measurements declines due to the distributed circuit parameters such as parasitic capacitance, on-resistance of switch and the series inductance, with a more significant impact on larger impedance. To overcome this limitation, this paper proposes an approach to enhance the precision of impedance measurement through self-identification of distributed parameter.<i>Approach.</i>Firstly, the distributed circuit parameters are identified via correction measurements of precision resistances in the frequency range from 5 kHz to 1 MHz; then, the circuit is accurately modeled; finally, transfer impedance measurements during imaging process are corrected using the established circuit model.<i>Main results.</i>The distributed circuit parameter self-identification method was verified through a goodness-of-fit test, confirming the consistency between the model's predicted values and the actual values of the component. The test results indicate that at 1 MHz, the relative residuals follow a right-skewed distribution with an average value of 0.08%, which demonstrates high model accuracy. At 1 MHz, the relative error after correction for the 499 Ω precision resistor measurement is reduced by 12.01%, and for the 56 pF precision capacitor in parallel with 249 Ω, the relative error after correction is 0.46%.<i>Significance.</i>The proposed method can extend the frequency range of EIT and other impedance technologies from below 200 kHz to up to 1 MHz, while ensuring good measurement accuracy.

Authors

Institutions

Publication Details

Journal
Physiological Measurement
Published
2025-02-24
DOI
https://doi.org/10.1088/1361-6579/adb9b4
Citations
1
Primary Topic
Electrical and Bioimpedance Tomography
Type
article
Field-Weighted Citation Impact
0.73
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Enhancing the precision of impedance measurement from 5 kHz to 1 MHz through self-identification of parasitic parameters

Zeyi Jiang, Sirui Qiao, Yixin Ma, Yi She et al.
1 citations
Physiological Measurement
Electrical and Bioimpedance Tomography
0.73
article

Enhancing the precision of impedance measurement from 5 kHz to 1 MHz through self-identification of parasitic parameters

Zeyi Jiang, Sirui Qiao, Yixin Ma, Yi She, Qin Liu
article en
1 citations

Abstract

<i>Objective.</i>Electrical impedance tomography (EIT) generates cross-sectional images through non-invasive impedance measurements from surface electrodes. While impedance above 200 kHz can reveal intracellular properties, most existing EIT images are published at frequencies below 200 kHz. When frequencies exceed 200 kHz, the accuracy of impedance measurements declines due to the distributed circuit parameters such as parasitic capacitance, on-resistance of switch and the series inductance, with a more significant impact on larger impedance. To overcome this limitation, this paper proposes an approach to enhance the precision of impedance measurement through self-identification of distributed parameter.<i>Approach.</i>Firstly, the distributed circuit parameters are identified via correction measurements of precision resistances in the frequency range from 5 kHz to 1 MHz; then, the circuit is accurately modeled; finally, transfer impedance measurements during imaging process are corrected using the established circuit model.<i>Main results.</i>The distributed circuit parameter self-identification method was verified through a goodness-of-fit test, confirming the consistency between the model's predicted values and the actual values of the component. The test results indicate that at 1 MHz, the relative residuals follow a right-skewed distribution with an average value of 0.08%, which demonstrates high model accuracy. At 1 MHz, the relative error after correction for the 499 Ω precision resistor measurement is reduced by 12.01%, and for the 56 pF precision capacitor in parallel with 249 Ω, the relative error after correction is 0.46%.<i>Significance.</i>The proposed method can extend the frequency range of EIT and other impedance technologies from below 200 kHz to up to 1 MHz, while ensuring good measurement accuracy.

Physiological MeasurementVol. 46(3)
Shanghai Jiao Tong University (CN)
Openalex Percentile: Top 31%
Electrical and Bioimpedance Tomography
0.73
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.