Fatigue test evaluation of a customised humeral component for an instrumented total elbow prosthesis and strain validation study

The survival rate of total elbow arthroplasty (TEA) is negatively impacted by the lack of available data on elbow biomechanics. This study developed a modified humeral component for TEA that is purposed to be instrumented to generate real-time 6 degrees of freedom (d.o.f) force and moment data during activities of daily living (ADL). The objectives are twofold: (1) to assess the safety of the modified humeral component under peak anticipated loads in fatigue, and (2) verify the strains measured under physiological loads with strains modelled using finite element analysis (FEA). Four modified titanium alloy humeral components were welded, and fatigue tested at 5 Hz for 5 million cycles under a compressive load of 700 N corresponding to moderate ADL. The strains were measured using triaxial 350 Ω rectangular rosette (45°) strain gauges bonded to three specific locations on the humeral component confirmed through an FE study. The four welded humeral components successfully withstood fatigue conditions and did not deform. The measured and modelled principal strains were confirmed to be highest at the external wall of the lateral cavity, with a percentage difference of <10 %.

Authors

Institutions

Publication Details

Journal
Medical Engineering & Physics
Published
2025-02-24
DOI
https://doi.org/10.1016/j.medengphy.2025.104311
Primary Topic
Shoulder Injury and Treatment
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Fatigue test evaluation of a customised humeral component for an instrumented total elbow prosthesis and strain validation study

Stephen Taylor, Simon Lambert, Marim Basiouny, Chin Kuenfoo
Medical Engineering & Physics
Shoulder Injury and Treatment
article

Fatigue test evaluation of a customised humeral component for an instrumented total elbow prosthesis and strain validation study

Stephen Taylor, Simon Lambert, Marim Basiouny, Chin Kuenfoo
article en

Abstract

The survival rate of total elbow arthroplasty (TEA) is negatively impacted by the lack of available data on elbow biomechanics. This study developed a modified humeral component for TEA that is purposed to be instrumented to generate real-time 6 degrees of freedom (d.o.f) force and moment data during activities of daily living (ADL). The objectives are twofold: (1) to assess the safety of the modified humeral component under peak anticipated loads in fatigue, and (2) verify the strains measured under physiological loads with strains modelled using finite element analysis (FEA). Four modified titanium alloy humeral components were welded, and fatigue tested at 5 Hz for 5 million cycles under a compressive load of 700 N corresponding to moderate ADL. The strains were measured using triaxial 350 Ω rectangular rosette (45°) strain gauges bonded to three specific locations on the humeral component confirmed through an FE study. The four welded humeral components successfully withstood fatigue conditions and did not deform. The measured and modelled principal strains were confirmed to be highest at the external wall of the lateral cavity, with a percentage difference of <10 %.

Medical Engineering & PhysicsVol. 138(1)
University College Hospital (GB), Barking, Havering And Redbridge University Hospitals NHS Trust (GB), Royal National Orthopaedic Hospital (GB), University College London (GB)
Zero hunger
Openalex Percentile: Top 94%
Shoulder Injury and Treatment
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.