Effect of Time Delay and Syringe Surface Area on Partial Pressure of Oxygen and Oxygen Saturation of Arterial and Venous Blood Gas Samples. observational study

Background: Many medical disorders require blood gas analysis, especially in severely ill patients who undergo venous or arterial blood gas testing. However, resource constraints and blood gas analyzer shortages can delay sample analysis in some hospitals. Objective: This study aimed to evaluate the effects of time delay and syringe surface area on oxygen tension (PaO2), oxyhemoglobin saturation (SaO2), acidity (pH), and carbon dioxide tension (PaCO2) in arterial and venous blood gas samples Patients and methods: This observational research included 45 ICU and intraoperative patients. Both arterial and central venous samples were collected and divided into five parts: 1 ml in a heparinized syringe for immediate analysis and 1 ml in a 3 ml plastic and capillary insulin syringe for delayed analysis (after 30 minutes and 1 hour). To analyze arterial and venous samples simultaneously, all samples were held at ambient temperatures during the waiting period. Therefore, each patient had 10 samples tested for the final assessment. Results: There was a wide range of limits of agreement for oxygen saturation, PaO2, and PaCO2. The mean baseline PaO2 for arterial samples was 159.9 ± 113.6 mmHg. The mean bias (limits of agreement) for the 30-minute and 60-minute values from baseline in the 3 ml syringe were 21 (-99 to 142) and 38 (-112 to 188), respectively. The mean baseline PaCO2 for arterial samples was 36 ± 7.6 mmHg, with the mean bias (limits of agreement) for the 30-minute and 60-minute values from baseline being -0.4 (-4.8 to 4.1) and -3.8 (-11.8 to 4.1), respectively, in the 3 ml syringe. Conclusion: Delayed arterial and venous samples stored at room temperature yield different results compared to those analyzed immediately, with significant changes detectable as early as 30 minutes. In clinical practice, using plastic syringes is more convenient and safer.

Authors

Institutions

Publication Details

Journal
International Journal of Medical Arts
Published
2025-02-22
DOI
https://doi.org/10.21608/ijma.2025.353227.2123
Primary Topic
Hemodynamic Monitoring and Therapy
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Effect of Time Delay and Syringe Surface Area on Partial Pressure of Oxygen and Oxygen Saturation of Arterial and Venous Blood Gas Samples. observational study

Dalia S. Shaker, Ahmed Lotfy, Adel Ali Hassan, M Hamid et al.
International Journal of Medical Arts
Hemodynamic Monitoring and Therapy
article

Effect of Time Delay and Syringe Surface Area on Partial Pressure of Oxygen and Oxygen Saturation of Arterial and Venous Blood Gas Samples. observational study

Dalia S. Shaker, Ahmed Lotfy, Adel Ali Hassan, M Hamid, H. E. A. Bakr, Amina Aboul Ela Elew
article en

Abstract

Background: Many medical disorders require blood gas analysis, especially in severely ill patients who undergo venous or arterial blood gas testing. However, resource constraints and blood gas analyzer shortages can delay sample analysis in some hospitals. Objective: This study aimed to evaluate the effects of time delay and syringe surface area on oxygen tension (PaO2), oxyhemoglobin saturation (SaO2), acidity (pH), and carbon dioxide tension (PaCO2) in arterial and venous blood gas samples Patients and methods: This observational research included 45 ICU and intraoperative patients. Both arterial and central venous samples were collected and divided into five parts: 1 ml in a heparinized syringe for immediate analysis and 1 ml in a 3 ml plastic and capillary insulin syringe for delayed analysis (after 30 minutes and 1 hour). To analyze arterial and venous samples simultaneously, all samples were held at ambient temperatures during the waiting period. Therefore, each patient had 10 samples tested for the final assessment. Results: There was a wide range of limits of agreement for oxygen saturation, PaO2, and PaCO2. The mean baseline PaO2 for arterial samples was 159.9 ± 113.6 mmHg. The mean bias (limits of agreement) for the 30-minute and 60-minute values from baseline in the 3 ml syringe were 21 (-99 to 142) and 38 (-112 to 188), respectively. The mean baseline PaCO2 for arterial samples was 36 ± 7.6 mmHg, with the mean bias (limits of agreement) for the 30-minute and 60-minute values from baseline being -0.4 (-4.8 to 4.1) and -3.8 (-11.8 to 4.1), respectively, in the 3 ml syringe. Conclusion: Delayed arterial and venous samples stored at room temperature yield different results compared to those analyzed immediately, with significant changes detectable as early as 30 minutes. In clinical practice, using plastic syringes is more convenient and safer.

International Journal of Medical ArtsVol. 0(0)
Cairo University (EG), Benha University (EG), Pain and Rehabilitation Medicine (US), Port Said University (EG)
Openalex Percentile: Top 94%
Hemodynamic Monitoring and Therapy
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.