Finite Element Analysis of Hydroxyapatite-Coated Titanium and Steel Hip Implants: Impacts on Stress Redistribution

This stress redistribution was consistent across both material types, highlighting the coatings' role in enhancing the load-bearing capacity of the implants. Furthermore, titanium implants exhibited lower stress concentrations compared to steel implants, confirming titanium's superior mechanical properties and biocompatibility. These findings suggest that the combination of titanium implants with HA coatings can substantially improve implant durability and performance, providing critical insights for optimizing hip implant designs to enhance patient outcomes. This study employs the Finite Element Method (FEM) to analyse and to mitigate stresses on hip implants, focusing specifically on the impact of Hydroxyapatite (HA) coatings on stress distribution. We examined both steel and titanium implants to assess the influence of material properties on stress patterns within the implant components. Our results demonstrated that HA coatings effectively shifted peak stress concentrations from the implant stem to the coating itself, leading to a significant reduction in overall stress levels. Specifically, the maximum stress in the steel stem without coating (model 1) decreased from 140.6 MPa to 66.1 MPa with the addition of the HA coating (model 2). Similarly, the maximum stress in the Ti-6Al-4V stem without coating (model 3) reduced from 96.9 MPa to 51.9 MPa with the coating (model 4). This stress redistribution was consistent across both material types, highlighting the coatings' role in enhancing the load-bearing capacity of the implants. Furthermore, titanium implants exhibited lower stress concentrations compared to steel implants, confirming titanium's superior mechanical properties and biocompatibility. These findings suggest that the combination of titanium implants with HA coatings can substantially improve implant durability and performance, providing critical insights for optimizing hip implant designs to enhance patient outcomes.

Authors

Publication Details

Journal
Acta Mechanica Slovaca
Published
2024-12-01
DOI
https://doi.org/10.21496/ams.2024.029
Citations
2
Primary Topic
Orthopaedic implants and arthroplasty
Type
article
Field-Weighted Citation Impact
1.15
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Finite Element Analysis of Hydroxyapatite-Coated Titanium and Steel Hip Implants: Impacts on Stress Redistribution

Atef Hamada, Abdelkader Benkhettou, Ismail Boudjemaa, Omar Khatir et al.
2 citations
Acta Mechanica Slovaca
Orthopaedic implants and arthroplasty
1.15
article

Finite Element Analysis of Hydroxyapatite-Coated Titanium and Steel Hip Implants: Impacts on Stress Redistribution

Atef Hamada, Abdelkader Benkhettou, Ismail Boudjemaa, Omar Khatir, Abderahmene Sahli
article en
2 citations

Abstract

This stress redistribution was consistent across both material types, highlighting the coatings' role in enhancing the load-bearing capacity of the implants. Furthermore, titanium implants exhibited lower stress concentrations compared to steel implants, confirming titanium's superior mechanical properties and biocompatibility. These findings suggest that the combination of titanium implants with HA coatings can substantially improve implant durability and performance, providing critical insights for optimizing hip implant designs to enhance patient outcomes. This study employs the Finite Element Method (FEM) to analyse and to mitigate stresses on hip implants, focusing specifically on the impact of Hydroxyapatite (HA) coatings on stress distribution. We examined both steel and titanium implants to assess the influence of material properties on stress patterns within the implant components. Our results demonstrated that HA coatings effectively shifted peak stress concentrations from the implant stem to the coating itself, leading to a significant reduction in overall stress levels. Specifically, the maximum stress in the steel stem without coating (model 1) decreased from 140.6 MPa to 66.1 MPa with the addition of the HA coating (model 2). Similarly, the maximum stress in the Ti-6Al-4V stem without coating (model 3) reduced from 96.9 MPa to 51.9 MPa with the coating (model 4). This stress redistribution was consistent across both material types, highlighting the coatings' role in enhancing the load-bearing capacity of the implants. Furthermore, titanium implants exhibited lower stress concentrations compared to steel implants, confirming titanium's superior mechanical properties and biocompatibility. These findings suggest that the combination of titanium implants with HA coatings can substantially improve implant durability and performance, providing critical insights for optimizing hip implant designs to enhance patient outcomes.

Acta Mechanica SlovacaVol. 28(4)
Climate action
Openalex Percentile: Top 18%
Orthopaedic implants and arthroplasty
1.15
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Finite Element Analysis of Hydroxyapatite-Coated Titanium and Steel Hip Implants: Impacts on Stress Redistribution — Atef Hamada, Abdelkader Benkhettou, et al. · Acta Mechanica Slovaca (2024) | TGRS Research Map | TGRS