Mesenteric Ischemia During Intra-Aortic Balloon Counterpulsation Therapy: Case Series
Intra-aortic balloon pumps (IABPs) are commonly used circulatory support devices. Mesenteric ischemia is a serious and poorly studied complication of IABP use. Here, we examine six cases of mesenteric ischemia associated with IABPs from a single urban academic medical center. We examine patient, device, and radiographic characteristics, and hypothesize several mechanisms behind IABP-associated mesenteric ischemia including anatomic-to-device length mismatch, atheroembolism, and mechanical obstruction. We highlight the high morbidity and mortality of such complication and the need for future studies to examine risk factors that may predispose IABP-treated patients to the development of mesenteric ischemia.
Authors
- Navin Kumar Kapur (ORCID: https://orcid.org/0000-0002-8302-6796)
- Haval Chweich (ORCID: https://orcid.org/0000-0001-6513-9220)
- Kevin John (ORCID: https://orcid.org/0000-0003-3382-0294)
- Alyssa Wohlfahrt (ORCID: https://orcid.org/0000-0003-4412-3275)
- Michael S. Kiernan (ORCID: https://orcid.org/0000-0003-4026-6383)
- Masashi Kawabori (ORCID: https://orcid.org/0000-0002-3580-5664)
- Alex Ablavsky
Institutions
- Tufts University (US)
- Tufts Medical Center (US)
- Newton Wellesley Hospital (US)
- Mass General Brigham (US)
Publication Details
- Journal
- ASAIO Journal
- Published
- 2025-02-24
- DOI
- https://doi.org/10.1097/mat.0000000000002397
- Primary Topic
- Mechanical Circulatory Support Devices
- Type
- article
- Field-Weighted Citation Impact
- 0.00