Human Blood Investigation of Ta‐Coated 316L Stainless Steel to Evaluate Corrosion and Wear Behaviour Against Other Artificial Fluids for Orthopaedic Implants

ABSTRACT Stainless steel (316L) is widely used in biomedical applications, but concerns persist regarding its long‐term surface properties, particularly in varying patient environments over time. This study aims to enhance the surface properties of 316L stainless steel through a 1.504 μm tantalum (Ta) coating applied via DC magnetron sputtering. The corrosion and wear behaviours were evaluated in different mediums, including human blood (HB), artificial saliva (AS), Ringer's solution (RS), and simulated body fluid (SBF). The Ta‐coated 316L SS exhibited a shift toward nobler corrosion behaviour, with a corrosion rate of 0.146 × 10 −2 mm/year in human blood, compared to the uncoated 316L SS, which showed corrosion rates of 0.325, 0.556, 0.779, and 1.425 × 10 −2 mm/year in HB, RS, AS, and SBF, respectively. Electrochemical impedance spectroscopy (EIS) results indicated higher corrosion resistance of Ta‐coated 316L SS in human blood than in other fluids. Potentiodynamic polarisation (PDP) results reveal that corrosion rate and current increased as the fluid environment shifted from human blood to simulated body fluid. Wear tests further demonstrated the superior wear resistance of the Ta‐coated 316L SS, with a minimum wear rate of 0.086 × 10 −5 mm³/Nm, compared to the uncoated 316L SS, which exhibited a maximum wear rate of 0.25 × 10 −5 mm³/Nm. The microstructural analysis confirmed significant improvements in the surface properties of Ta‐coated 316L SS, particularly in the human blood environment.

Authors

Institutions

Publication Details

Journal
Materials and Corrosion
Published
2025-02-23
DOI
https://doi.org/10.1002/maco.202414727
Primary Topic
Orthopaedic implants and arthroplasty
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Human Blood Investigation of Ta‐Coated 316L Stainless Steel to Evaluate Corrosion and Wear Behaviour Against Other Artificial Fluids for Orthopaedic Implants

C.K. Behera, Abhijit Mandal, Dileep Pathote, Rakesh Kumar Gautam et al.
2 links
Materials and Corrosion
Orthopaedic implants and arthroplasty
article

Human Blood Investigation of Ta‐Coated 316L Stainless Steel to Evaluate Corrosion and Wear Behaviour Against Other Artificial Fluids for Orthopaedic Implants

C.K. Behera, Abhijit Mandal, Dileep Pathote, Rakesh Kumar Gautam, Ashwani Ranjan, Sandeep Kumar
article en

Abstract

ABSTRACT Stainless steel (316L) is widely used in biomedical applications, but concerns persist regarding its long‐term surface properties, particularly in varying patient environments over time. This study aims to enhance the surface properties of 316L stainless steel through a 1.504 μm tantalum (Ta) coating applied via DC magnetron sputtering. The corrosion and wear behaviours were evaluated in different mediums, including human blood (HB), artificial saliva (AS), Ringer's solution (RS), and simulated body fluid (SBF). The Ta‐coated 316L SS exhibited a shift toward nobler corrosion behaviour, with a corrosion rate of 0.146 × 10 −2 mm/year in human blood, compared to the uncoated 316L SS, which showed corrosion rates of 0.325, 0.556, 0.779, and 1.425 × 10 −2 mm/year in HB, RS, AS, and SBF, respectively. Electrochemical impedance spectroscopy (EIS) results indicated higher corrosion resistance of Ta‐coated 316L SS in human blood than in other fluids. Potentiodynamic polarisation (PDP) results reveal that corrosion rate and current increased as the fluid environment shifted from human blood to simulated body fluid. Wear tests further demonstrated the superior wear resistance of the Ta‐coated 316L SS, with a minimum wear rate of 0.086 × 10 −5 mm³/Nm, compared to the uncoated 316L SS, which exhibited a maximum wear rate of 0.25 × 10 −5 mm³/Nm. The microstructural analysis confirmed significant improvements in the surface properties of Ta‐coated 316L SS, particularly in the human blood environment.

Materials and CorrosionVol. 76(7)
Indian Institute of Technology BHU (IN), Banaras Hindu University (IN)
Openalex Percentile: Top 94%
Orthopaedic implants and arthroplasty
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.