Characterization, commissioning, and clinical evaluation of a commercial BeO optically stimulated luminescence (OSL) system

This article investigates the performance of a commercial BeO optically stimulated luminescent (OSL) dosimetry system (myOSLchip, RadPro GmbH International, Remscheid, Germany) through the application of the commissioning framework for luminescent dosimeters as described in the American Association of Physicists in Medicine Task Group 191 (AAPM TG191) report. Initial clinical experiences and dosimetric results are also presented. The following properties of the system were characterized: linearity correction factors ranged from -0.5% to +3% for dose levels spanning 0.1 to 20 Gy. Beam quality correction factors (relative to 6 MV) ranged from -4.5% (2.5FFF) to +4.5% (15MV) for photon beams and +1.9% (6 MeV) to +4.3% (20 MeV) for electron beams. An average (µ) signal loss per reading of -2.13% ± 0.20% was measured, however greater signal loss was observed in the first reading (µ = -2.6% ± 0.46%). An initial decline in individual element sensitivity relative to baseline was observed from 0-15 Gy cumulative dose (µ = -1.98% ± 0.55%), with negligible further deterioration from 15-32 Gy (µ = -2.38% ± 0.85%). Post-irradiation, there was a transient OSL signal which faded with a half-life of 1.8 min; this signal enhancement was +5% at 5 min post-irradiation and +1% at 15 min relative to 24 h. Dosimeter response was not dependent on average dose rate in the range of 100-2500 MU/min. With respect to clinical testing, equal or superior performance compared with aluminum oxide OSLs (nanoDots) is shown for a range of clinical techniques and modalities including TSET, TBI, en-face electrons, and pacemaker/out-of-field measurements. The feasibility of myOSLchip to serve as a primary clinical in vivo dosimetry system and direct replacement for Landauer's microStar system is demonstrated.

Authors

Institutions

Publication Details

Journal
Journal of Applied Clinical Medical Physics
Published
2025-02-22
DOI
https://doi.org/10.1002/acm2.70057
Citations
7
Primary Topic
Advanced Radiotherapy Techniques
Type
article
Field-Weighted Citation Impact
13.18
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Characterization, commissioning, and clinical evaluation of a commercial BeO optically stimulated luminescence (OSL) system

Xinyi Li, Qiuwen Wu, Brett G. Erickson, Sua Yoo et al.
7 citations
Journal of Applied Clinical Medical Physics
Advanced Radiotherapy Techniques
13.18
article

Characterization, commissioning, and clinical evaluation of a commercial BeO optically stimulated luminescence (OSL) system

Xinyi Li, Qiuwen Wu, Brett G. Erickson, Sua Yoo, J. Kowalski
article en
7 citations

Abstract

This article investigates the performance of a commercial BeO optically stimulated luminescent (OSL) dosimetry system (myOSLchip, RadPro GmbH International, Remscheid, Germany) through the application of the commissioning framework for luminescent dosimeters as described in the American Association of Physicists in Medicine Task Group 191 (AAPM TG191) report. Initial clinical experiences and dosimetric results are also presented. The following properties of the system were characterized: linearity correction factors ranged from -0.5% to +3% for dose levels spanning 0.1 to 20 Gy. Beam quality correction factors (relative to 6 MV) ranged from -4.5% (2.5FFF) to +4.5% (15MV) for photon beams and +1.9% (6 MeV) to +4.3% (20 MeV) for electron beams. An average (µ) signal loss per reading of -2.13% ± 0.20% was measured, however greater signal loss was observed in the first reading (µ = -2.6% ± 0.46%). An initial decline in individual element sensitivity relative to baseline was observed from 0-15 Gy cumulative dose (µ = -1.98% ± 0.55%), with negligible further deterioration from 15-32 Gy (µ = -2.38% ± 0.85%). Post-irradiation, there was a transient OSL signal which faded with a half-life of 1.8 min; this signal enhancement was +5% at 5 min post-irradiation and +1% at 15 min relative to 24 h. Dosimeter response was not dependent on average dose rate in the range of 100-2500 MU/min. With respect to clinical testing, equal or superior performance compared with aluminum oxide OSLs (nanoDots) is shown for a range of clinical techniques and modalities including TSET, TBI, en-face electrons, and pacemaker/out-of-field measurements. The feasibility of myOSLchip to serve as a primary clinical in vivo dosimetry system and direct replacement for Landauer's microStar system is demonstrated.

Journal of Applied Clinical Medical PhysicsVol. 26(4)
Mayo Clinic (US), Duke Medical Center (US)
Openalex Percentile: Top 1%
Advanced Radiotherapy Techniques
13.18
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.