Determination of Cr (VI) released by leather samples using smartphone-based colorimetry for on-site determination and miniaturization as greener preparation strategy

• A miniaturized method for detecting Cr (VI) released by leather samples. • Smartphone-based analyses achieve LOD = 0.0202 µg mL -1 and LOQ = 0.0612 µg mL -1 . • Reduced environmental impacts compared to the reference method. In this work a lab-on-a-smartphone platform for on-site and rapid assessment of Cr (VI) released by leather samples has been proposed. In comparison with the reference method (BS EN ISO 17,075–1:2017) the newly developed one permits to reduce the amount of extractant solution up to 10 mL (100 mL for the reference method) through a miniaturization step. The released Cr (VI) was determined after the reaction with 1,5-diphenylcarbazide using a smartphone equipped with Photometrix®. This approach achieves LOD = 0.0202 µg mL -1 and LOQ = 0.0612 µg mL -1 very closed to the reference method which allows the adoption of this miniaturized method for analysis of real samples. The analyses were carried out on Cr-tanned leather samples with encouraging results favouring the on-site determination which will be further explored as quality control test in the leather industry. The developed method permits to reduce environmental impacts, determined by using both AGREE and AGREEprep, related to Cr (VI) determination in terms of produced waste and energy consumption allowing the process of an higher number of samples as a function of time.

Authors

Institutions

Publication Details

Journal
Green Analytical Chemistry
Published
2025-02-22
DOI
https://doi.org/10.1016/j.greeac.2025.100232
Citations
2
Primary Topic
Analytical chemistry methods development
Type
article
Field-Weighted Citation Impact
1.50

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Determination of Cr (VI) released by leather samples using smartphone-based colorimetry for on-site determination and miniaturization as greener preparation strategy

Raffaele Cucciniello, Adriano Intiso, Prisco Prete, Sonia Acocella
2 citations
Green Analytical Chemistry
Analytical chemistry methods development
1.50
article

Determination of Cr (VI) released by leather samples using smartphone-based colorimetry for on-site determination and miniaturization as greener preparation strategy

Raffaele Cucciniello, Adriano Intiso, Prisco Prete, Sonia Acocella
article en
2 citations

Abstract

• A miniaturized method for detecting Cr (VI) released by leather samples. • Smartphone-based analyses achieve LOD = 0.0202 µg mL -1 and LOQ = 0.0612 µg mL -1 . • Reduced environmental impacts compared to the reference method. In this work a lab-on-a-smartphone platform for on-site and rapid assessment of Cr (VI) released by leather samples has been proposed. In comparison with the reference method (BS EN ISO 17,075–1:2017) the newly developed one permits to reduce the amount of extractant solution up to 10 mL (100 mL for the reference method) through a miniaturization step. The released Cr (VI) was determined after the reaction with 1,5-diphenylcarbazide using a smartphone equipped with Photometrix®. This approach achieves LOD = 0.0202 µg mL -1 and LOQ = 0.0612 µg mL -1 very closed to the reference method which allows the adoption of this miniaturized method for analysis of real samples. The analyses were carried out on Cr-tanned leather samples with encouraging results favouring the on-site determination which will be further explored as quality control test in the leather industry. The developed method permits to reduce environmental impacts, determined by using both AGREE and AGREEprep, related to Cr (VI) determination in terms of produced waste and energy consumption allowing the process of an higher number of samples as a function of time.

Green Analytical ChemistryVol. 12
Istituto di Scienza e Tecnologie dell'Informazione "Alessandro Faedo" (IT), University of Salerno (IT), Accademia Nazionale dei Lincei (IT)
Università degli Studi di Salerno
Openalex Percentile: Top 22%
Analytical chemistry methods development
1.50
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Determination of Cr (VI) released by leather samples using smartphone-based colorimetry for on-site determination and miniaturization as greener preparation strategy — Raffaele Cucciniello, Adriano Intiso, et al. · Green Analytical Chemistry (2025) | TGRS Research Map | TGRS