Is thoron a problem in radon measurements with NRPB/SSI passive radon detectors—experimental study

The UKHSA radon detectors with polyallyldiglycol carbonate sensing material produced from 2010 onward were used to confirm that they are purely radon detectors only with negligible interference from thoron presence. The statistical analysis of results, after following standard detector processing procedures, showed that such devices can only detect 2.4% ± 0.3% of the total thoron activity concentration compared to the calibrated reference device. It was also demonstrated that thoron, unlike radon, can only travel a very short distance from the source to be effectively measured. Based on the above, it has been concluded that interference of thoron in radon measurement is negligible (within the statistical measurement error), and hence can be ignored in standard indoor and outdoor measurements.

Authors

Institutions

Publication Details

Journal
Radiation Protection Dosimetry
Published
2025-02-21
DOI
https://doi.org/10.1093/rpd/ncaf012
Citations
1
Primary Topic
Radioactivity and Radon Measurements
Type
article
Field-Weighted Citation Impact
1.54
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Is thoron a problem in radon measurements with NRPB/SSI passive radon detectors—experimental study

Z. Daraktchieva, K. Mitev, S. Georgiev, Jaroslaw M. Wasikiewicz et al.
1 citations
Radiation Protection Dosimetry
Radioactivity and Radon Measurements
1.54
article

Is thoron a problem in radon measurements with NRPB/SSI passive radon detectors—experimental study

Z. Daraktchieva, K. Mitev, S. Georgiev, Jaroslaw M. Wasikiewicz, Ì. Dimitrova, Z.-F. Ibrahimi
article en
1 citations

Abstract

The UKHSA radon detectors with polyallyldiglycol carbonate sensing material produced from 2010 onward were used to confirm that they are purely radon detectors only with negligible interference from thoron presence. The statistical analysis of results, after following standard detector processing procedures, showed that such devices can only detect 2.4% ± 0.3% of the total thoron activity concentration compared to the calibrated reference device. It was also demonstrated that thoron, unlike radon, can only travel a very short distance from the source to be effectively measured. Based on the above, it has been concluded that interference of thoron in radon measurement is negligible (within the statistical measurement error), and hence can be ignored in standard indoor and outdoor measurements.

Radiation Protection DosimetryVol. 201(4)
Devon History Society (GB), Sofia University "St. Kliment Ohridski" (BG)
Openalex Percentile: Top 22%
Radioactivity and Radon Measurements
1.54
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.