Evaluation of Using an Octavius 4D Measuring System for Patient-Specific VMAT Quality Assurance

Background: Quality assurance (QA) programs are designed to improve the quality and safety of radiation treatments, including patient-specific QA (PSQA). The objective of this study was to investigate the conditions in which pretreatment PSQA is performed, to evaluate the root cause of the implementation of more complex techniques, and to identify areas for potential improvement. Materials/Methods: The Octavius 4D (O4D) system accuracy was evaluated using an O4D homogeneous phantom for different field sizes. Tests of the system response to dose linearity, field sizes, and PDD differences were performed against calculated doses for a 6 MV photon beam. The pretreatment verification of 40 VMAT plans was performed using the PTW VeriSoft software (version 8.0.1) for local and global 3D gamma analysis. The reconstructed 3D dose was compared to the calculated dose using 2%/2 mm and 3%/3 mm, 20% of the low-dose threshold, and 95% of the gamma passing rate (%GP) tolerance level. The sensitivity of the O4D system in detecting VMAT delivery and setup errors has been investigated by measuring the variation in %GP values before and after the simulated errors. Results: The O4D system reported good agreement for linearity, field size, and PDD differences with TPS dose, being within ±2% tolerance. The output factors were consistent between the ionization chamber and the O4D detector down to a 4 × 4 cm2 field size with a maximum deviation less than 1%. The introduction of deliberate errors caused a decrease in %GP values. In most scenarios, the %GP value of the simulated errors was detected with 2%/2 mm. Conclusion: The results indicate that the O4D system is sensitive enough to detect delivery and setup errors with the restrictive global criterion of 2%/2 mm for routine pretreatment verification.

Authors

Institutions

Publication Details

Journal
Radiation
Published
2025-02-20
DOI
https://doi.org/10.3390/radiation5010009
Primary Topic
Advanced Radiotherapy Techniques
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Evaluation of Using an Octavius 4D Measuring System for Patient-Specific VMAT Quality Assurance

Yawo A.C. Fiagan, Ahmed Diakité, Dirk Verellen, Thierry Gevaert et al.
Radiation
Advanced Radiotherapy Techniques
article

Evaluation of Using an Octavius 4D Measuring System for Patient-Specific VMAT Quality Assurance

Yawo A.C. Fiagan, Ahmed Diakité, Dirk Verellen, Thierry Gevaert, Kodjo Joël Fabrice N’Guessan, Komlanvi Adjénou
article en

Abstract

Background: Quality assurance (QA) programs are designed to improve the quality and safety of radiation treatments, including patient-specific QA (PSQA). The objective of this study was to investigate the conditions in which pretreatment PSQA is performed, to evaluate the root cause of the implementation of more complex techniques, and to identify areas for potential improvement. Materials/Methods: The Octavius 4D (O4D) system accuracy was evaluated using an O4D homogeneous phantom for different field sizes. Tests of the system response to dose linearity, field sizes, and PDD differences were performed against calculated doses for a 6 MV photon beam. The pretreatment verification of 40 VMAT plans was performed using the PTW VeriSoft software (version 8.0.1) for local and global 3D gamma analysis. The reconstructed 3D dose was compared to the calculated dose using 2%/2 mm and 3%/3 mm, 20% of the low-dose threshold, and 95% of the gamma passing rate (%GP) tolerance level. The sensitivity of the O4D system in detecting VMAT delivery and setup errors has been investigated by measuring the variation in %GP values before and after the simulated errors. Results: The O4D system reported good agreement for linearity, field size, and PDD differences with TPS dose, being within ±2% tolerance. The output factors were consistent between the ionization chamber and the O4D detector down to a 4 × 4 cm2 field size with a maximum deviation less than 1%. The introduction of deliberate errors caused a decrease in %GP values. In most scenarios, the %GP value of the simulated errors was detected with 2%/2 mm. Conclusion: The results indicate that the O4D system is sensitive enough to detect delivery and setup errors with the restrictive global criterion of 2%/2 mm for routine pretreatment verification.

RadiationVol. 5(1)
Vrije Universiteit Brussel (BE), University of Antwerp (BE), Universitair Ziekenhuis Brussel (BE), Iridium Kankernetwerk (BE), University of Lomé (TG)
Openalex Percentile: Top 96%
Advanced Radiotherapy Techniques
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.