Disinfection of outpatient ophthalmic devices: a critique of “semicritical” designation

Ophthalmic devices used in outpatient clinics are undergoing scrutiny by regulatory agencies and hospital infection control groups. Applying a nearly century-old classification for disinfection and sterilization of reusable devices and instruments, they are faulting ophthalmologists for not using high-level disinfection (HLD) listed in newer manufacturer instructions for use. By this classification, A-scan and B-scan ultrasound probes, fundus and laser contact lenses, and gonioscopy lenses are grouped with other "semicritical" devices, such as gastrointestinal endoscopes, laryngoscope blades, and anorectal manometers, and therefore require sterilization or HLD. HLD, however, varies by geographic region, compatibility with material, and disinfectant availability. Because HLD necessitates turnover time, clinics must buy more devices to maintain patient volume. Therefore, manufacturers have no incentive to develop sustainable or reusable devices. Most important, however, none of the ophthalmic devices in question has been associated with spread of infection. In short, the mandate for HLD seems to be arbitrary, reactive to perception of possible infection risk from classification of the eye as a "mucous membrane," and devoid of evidence that HLD decreases risk of infection transmission relative to current cleaning and disinfection methods. In the manner that the Ophthalmic Instrument Cleaning and Sterilization Task Force Regulatory raised concerns that regulatory agencies arbitrarily were imposing costly infection control measures for intraocular ophthalmic surgery without evidence that they improved patient safety, a similar task force is needed to address regulatory HLD mandates for clinic devices used in tens of millions of patient encounters for decades without infection transmission.

Authors

Institutions

Publication Details

Journal
Journal of Cataract & Refractive Surgery
Published
2025-02-21
DOI
https://doi.org/10.1097/j.jcrs.0000000000001635
Citations
2
Primary Topic
Retinal and Optic Conditions
Type
article
Field-Weighted Citation Impact
3.41
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Disinfection of outpatient ophthalmic devices: a critique of “semicritical” designation

Irene C. Kuo
2 citations
Journal of Cataract & Refractive Surgery
Retinal and Optic Conditions
3.41
article

Disinfection of outpatient ophthalmic devices: a critique of “semicritical” designation

Irene C. Kuo
article en
2 citations

Abstract

Ophthalmic devices used in outpatient clinics are undergoing scrutiny by regulatory agencies and hospital infection control groups. Applying a nearly century-old classification for disinfection and sterilization of reusable devices and instruments, they are faulting ophthalmologists for not using high-level disinfection (HLD) listed in newer manufacturer instructions for use. By this classification, A-scan and B-scan ultrasound probes, fundus and laser contact lenses, and gonioscopy lenses are grouped with other "semicritical" devices, such as gastrointestinal endoscopes, laryngoscope blades, and anorectal manometers, and therefore require sterilization or HLD. HLD, however, varies by geographic region, compatibility with material, and disinfectant availability. Because HLD necessitates turnover time, clinics must buy more devices to maintain patient volume. Therefore, manufacturers have no incentive to develop sustainable or reusable devices. Most important, however, none of the ophthalmic devices in question has been associated with spread of infection. In short, the mandate for HLD seems to be arbitrary, reactive to perception of possible infection risk from classification of the eye as a "mucous membrane," and devoid of evidence that HLD decreases risk of infection transmission relative to current cleaning and disinfection methods. In the manner that the Ophthalmic Instrument Cleaning and Sterilization Task Force Regulatory raised concerns that regulatory agencies arbitrarily were imposing costly infection control measures for intraocular ophthalmic surgery without evidence that they improved patient safety, a similar task force is needed to address regulatory HLD mandates for clinic devices used in tens of millions of patient encounters for decades without infection transmission.

Journal of Cataract & Refractive SurgeryVol. 51(5)
Johns Hopkins University (US), Johns Hopkins Medicine (US)
Openalex Percentile: Top 10%
Retinal and Optic Conditions
3.41
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Disinfection of outpatient ophthalmic devices: a critique of “semicritical” designation — Irene C. Kuo · Journal of Cataract & Refractive Surgery (2025) | TGRS Research Map | TGRS