SAFIR-II: Performance Evaluation of a High-Rate Preclinical PET-MR System

SAFIR-II is a preclinical PET insert compatible with a Bruker BioSpec 70/30 magnetic resonance imaging (MRI) scanner. It was designed to acquire data at activities of up to 500 MBq, enabling truly simultaneous preclinical positron emission tomography magnetic resonance imaging for mice and rats using image acquisition times of as little as 5 s. We present a brief overview of the system’s design as well as the results of several performance evaluations. SAFIR-II features an axial field-of-view (FOV) of 145 mm, covered by lutetium-yttrium oxyorthosilicate crystals coupled to Hamamatsu silicon photomultiplier (SiPM) arrays. PETA8 application-specific integrated circuits are used to digitize the SiPM’s analog signals, and custom MR-compatible dc-dc converters condition the system’s internal voltages. The insert exhibits a coincidence timing resolution of 221-ps full width at half maximum (FWHM), a coincidence energy resolution of 12.1%, and a peak sensitivity of 3.89% observed following the NEMA-NU4 standard. It is capable of resolving 1.7-mm hot rods within a Derenzo phantom filled with <inline-formula xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink"> <tex-math notation="LaTeX">$^{18}{\mathrm { F}}$ </tex-math></inline-formula> and features a peak noise-equivalent count rate of 1.12 Mcps observed at an activity of 451 MBq using the NEMA rat-like phantom. We furthermore present an evaluation of the system’s image quality determined using a NEMA image quality phantom, an evaluation of its MRI-compatibility, as well as images from an initial in vivo measurement using a Sprague-Dawley rat injected with 283-MBq fluordesoxyglucose.

Authors

Publication Details

Journal
IEEE Transactions on Radiation and Plasma Medical Sciences
Published
2025-02-21
DOI
https://doi.org/10.1109/trpms.2025.3542994
Primary Topic
Medical Imaging Techniques and Applications
Type
article
Field-Weighted Citation Impact
0.00

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

SAFIR-II: Performance Evaluation of a High-Rate Preclinical PET-MR System

W. Lustermann, Matthias T. Wyss, Bruno Weber, Jan Debus et al.
IEEE Transactions on Radiation and Plasma Medical Sciences
Medical Imaging Techniques and Applications
article

SAFIR-II: Performance Evaluation of a High-Rate Preclinical PET-MR System

W. Lustermann, Matthias T. Wyss, Bruno Weber, Jan Debus, Afroditi Eleftheriou, Günther Dissertori
article en

Abstract

SAFIR-II is a preclinical PET insert compatible with a Bruker BioSpec 70/30 magnetic resonance imaging (MRI) scanner. It was designed to acquire data at activities of up to 500 MBq, enabling truly simultaneous preclinical positron emission tomography magnetic resonance imaging for mice and rats using image acquisition times of as little as 5 s. We present a brief overview of the system’s design as well as the results of several performance evaluations. SAFIR-II features an axial field-of-view (FOV) of 145 mm, covered by lutetium-yttrium oxyorthosilicate crystals coupled to Hamamatsu silicon photomultiplier (SiPM) arrays. PETA8 application-specific integrated circuits are used to digitize the SiPM’s analog signals, and custom MR-compatible dc-dc converters condition the system’s internal voltages. The insert exhibits a coincidence timing resolution of 221-ps full width at half maximum (FWHM), a coincidence energy resolution of 12.1%, and a peak sensitivity of 3.89% observed following the NEMA-NU4 standard. It is capable of resolving 1.7-mm hot rods within a Derenzo phantom filled with <inline-formula xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink"> <tex-math notation="LaTeX">$^{18}{\mathrm { F}}$ </tex-math></inline-formula> and features a peak noise-equivalent count rate of 1.12 Mcps observed at an activity of 451 MBq using the NEMA rat-like phantom. We furthermore present an evaluation of the system’s image quality determined using a NEMA image quality phantom, an evaluation of its MRI-compatibility, as well as images from an initial in vivo measurement using a Sprague-Dawley rat injected with 283-MBq fluordesoxyglucose.

IEEE Transactions on Radiation and Plasma Medical SciencesVol. 9(7)
ETH Zürich Foundation
Zero hunger
Openalex Percentile: Top 95%
Medical Imaging Techniques and Applications
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.