Alleviating the trade-off between coincidence time resolution and sensitivity using scalable TOF-DOI detectors

<i>Objective</i>. Coincidence time resolution (CTR) in time-of-flight positron emission tomography (TOF-PET) has significantly improved with advancements in scintillators, photodetectors, and readout electronics. Achieving a CTR of 100 ps remains challenging due to the need for sufficiently thick scintillators-typically 20 mm-to ensure adequate sensitivity because the photon transit time spread within these thick scintillators impedes achieving 100 ps CTR. Therefore thinner scintillators are preferable for CTR better than 100 ps. To address the trade-off between TOF capability and sensitivity, we propose a readout scheme of PET detectors.<i>Approach</i>. The proposed scheme utilizes two orthogonally stacked one-dimensional PET detectors, enabling the thickness of the scintillators to be reduced to approximately 13 mm without compromising sensitivity. This is achieved by stacking the detectors along the depth-of-interaction (DOI) axis of a PET scanner. We refer to this design as the cross-stacked detector, or xDetector. Furthermore, the xDetector inherently provides DOI information using the same readout scheme.<i>Main results</i>. Experimental evaluations demonstrated that the xDetector achieved the best CTR of 175 ps full width at half maximum (FWHM) and an energy resolution of 11% FWHM at 511 keV with 3 × 3 × 12.8 mm<sup>3</sup>lutetium oxyorthosilicate crystals, each coupled one-to-one with silicon photomultipliers. The CTRs are between the xDetector and reference detector with a single timing resolution of 111.2 ± 0.8 ps FWHM. In terms of<i>xy</i>-spatial resolution, the xDetector exhibited an asymmetric resolution due to its readout scheme: one resolution was defined by the 3.2 mm readout pitch, while the other was calculated using the center-of-gravity method.<i>Significance</i>. The xDetector effectively resolves the trade-off between TOF capability and sensitivity while offering scalability and DOI capability. By integrating state-of-the-art scintillators, photodetectors, and readout electronics with the xDetector scheme, achieving a CTR of 100 ps FWHM alongside high DOI resolution becomes a practical possibility.

Authors

Institutions

Publication Details

Journal
Physics in Medicine and Biology
Published
2025-02-21
DOI
https://doi.org/10.1088/1361-6560/adb936
Citations
1
Primary Topic
Medical Imaging Techniques and Applications
Type
article
Field-Weighted Citation Impact
1.36
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Alleviating the trade-off between coincidence time resolution and sensitivity using scalable TOF-DOI detectors

Yuya Onishi, Ryosuke Ota
1 citations
Physics in Medicine and Biology
Medical Imaging Techniques and Applications
1.36
article

Alleviating the trade-off between coincidence time resolution and sensitivity using scalable TOF-DOI detectors

Yuya Onishi, Ryosuke Ota
article en
1 citations

Abstract

<i>Objective</i>. Coincidence time resolution (CTR) in time-of-flight positron emission tomography (TOF-PET) has significantly improved with advancements in scintillators, photodetectors, and readout electronics. Achieving a CTR of 100 ps remains challenging due to the need for sufficiently thick scintillators-typically 20 mm-to ensure adequate sensitivity because the photon transit time spread within these thick scintillators impedes achieving 100 ps CTR. Therefore thinner scintillators are preferable for CTR better than 100 ps. To address the trade-off between TOF capability and sensitivity, we propose a readout scheme of PET detectors.<i>Approach</i>. The proposed scheme utilizes two orthogonally stacked one-dimensional PET detectors, enabling the thickness of the scintillators to be reduced to approximately 13 mm without compromising sensitivity. This is achieved by stacking the detectors along the depth-of-interaction (DOI) axis of a PET scanner. We refer to this design as the cross-stacked detector, or xDetector. Furthermore, the xDetector inherently provides DOI information using the same readout scheme.<i>Main results</i>. Experimental evaluations demonstrated that the xDetector achieved the best CTR of 175 ps full width at half maximum (FWHM) and an energy resolution of 11% FWHM at 511 keV with 3 × 3 × 12.8 mm<sup>3</sup>lutetium oxyorthosilicate crystals, each coupled one-to-one with silicon photomultipliers. The CTRs are between the xDetector and reference detector with a single timing resolution of 111.2 ± 0.8 ps FWHM. In terms of<i>xy</i>-spatial resolution, the xDetector exhibited an asymmetric resolution due to its readout scheme: one resolution was defined by the 3.2 mm readout pitch, while the other was calculated using the center-of-gravity method.<i>Significance</i>. The xDetector effectively resolves the trade-off between TOF capability and sensitivity while offering scalability and DOI capability. By integrating state-of-the-art scintillators, photodetectors, and readout electronics with the xDetector scheme, achieving a CTR of 100 ps FWHM alongside high DOI resolution becomes a practical possibility.

Physics in Medicine and BiologyVol. 70(6)
Hamamatsu Photonics (Japan) (JP)
Affordable and clean energy
Openalex Percentile: Top 23%
Medical Imaging Techniques and Applications
1.36
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.